The aim of the study is to confirm the safety and clinical effectiveness of Nipro ELISIO-15H and to compare its performance characteristics and hemocompatibility with those of the commercially available dialyzers in U.S.A., Gambro Polyflux Revaclear and Fresenius F160NRe. 10 stable patients will be dialyzed 3 times/week with each dialyzer over 3 weeks. Blood samples for several parameters will be collected during hemodialysis to assess the performance and hemocompatibility of the dialyzers under study. Data thus collected will be analyzed for comparison by standard statistical analysis.
Current patients receiving dialysis treatment for end stage renal disease and who meet the inclusion/exclusion criteria will be entered into the study. These patients will continue on their standard dialysis regimen as stated in the clinics Standard Operating Procedures. Each patient will be studied for three weeks with each week starting with a dialyzer of a different manufacture but of comparable characteristics. Solute clearances and removal will be determined at each of the three dialysis sessions during the week. During session 2 of each week additional specimens will be taken to analyze for hemocompatibility. At the end of the study all clearances and the hemocompatibility markers for each dialyzer will be compared to one another as well as to the manufactures stated characteristics. Blood for the clearance and removal calculations will be taken at time 0, 60 minutes and 240 minutes. Blood for the hemocompatibility determinations will be taken at time 0, 15, 30, 60 and 240 minutes. Flow rates, anticoagulant therapy, erythropoietin therapy, venous and arterial pressures will also be recorded.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
BASIC_SCIENCE
Masking
NONE
Enrollment
11
Each patient to undergo 3 successive hemodialysis/week with this dialyzer.
Each patient to undergo 3 successive hemodialysis/week with this dialyzer.
Each patient to undergo 3 successive hemodialysis/week with this dialyzer.
Dialysis Clinic Incorporated
Columbia, Missouri, United States
Clearance and Solute Removal Characteristics of Dialyzers
Clearances of small, middle and large molecular weight solutes will be measured at 60 minute time point utilizing urea, creatinine, beta 2 microglobulin, phosphate and myoglobin. Dialyzer clearance is calculated as follows: 60 minute clearance:(((\[S\]60'Art - \[S\]60'Ven) / \[S\]60'Art) \* Qb) where \[S\] = solute concentration; Qb = blood flow rate. S
Time frame: During each of the three weekly dialysis sessions measurements of urea were made at the 60 minute time point for each patient in the study. At the end of the study all measurements were summed for each of the three dialyzers.
Hemocompatibility as Shown by C5a for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers of Dialyzers
The molecules monitored will be C5a, thrombin/anti-thrombin complex and a complete blood count with platelets.
Time frame: Once a week during the second of the three weekly sessions specimens will be taken at time points 0, 15, 30, 60 and 240 minutes. Measurements, all patients, were summed for each dialyzer type by time period.
Hemocompatibility as Shown by Thrombin/Antithrombin Complex for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers.
Measure of the % change from Baseline for the hemocompatibility marker Thrombin/Antithrombin complex for the Optiflux, Revaclear, and ELISIO dialyzers.
Time frame: Once a week during the second of the three weekly sessions specimens will be taken at the 60 and 240 minutes timepoints. Measurements for all patients were summed for each dialyzer type by time period for analysis..
Hemocompatibility as Shown by Hematocrit for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers
Measure of the % change from Baseline for the hemocompatibility marker Hematocrit for the Optiflux, Revaclear, and ELISIO dialyzers.
Time frame: Once a week during the second of three weekly sessions samples will be taken at the 0, 15, 30, 60, and 240 minute timepoints. Measurements for all patients were summed for each dialyzer type by time period for analysis.
Hemocompatibility as Shown by Hemoglobin for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers
Measure of the % change from Baseline for the hemocompatibility marker Hemoglobin for the Optiflux, Revaclear, and ELISIO dialyzers.
Time frame: Once a week during the second of the three weekly sessions samples will be taken at the 0, 15, 30, 60 and 240 minutes time points. Measurements for all patients were summed for each dialyzer type and time point for analysis.
Hemocompatibility as Shown by White Cell Count for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers
Measure of the % change from Baseline for the hemocompatibility marker White Cell Count for the Optiflux, Revaclear, and ELISIO dialyzers.
Time frame: Once a week during the second of the three weekly sessions samples will be taken at the 0, 15, 30, 60 and 240 minutes time periods. All measurements for all patients were summed for each dialyzer type and time point.
Hemocompatibility as Shown by Platelets for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers
Measure of the % change from Baseline for the hemocompatibility marker Platelets for the Optiflux, Revaclear, and ELISIO dialyzers.
Time frame: Once a week during the second of the three weekly sessions samples will be taken at the 0, 15, 30, 60 and 240 minutes time periods. All measurements for all patients were summed for each dialyzer type and time point.