The first purpose of this study is to test a new computer program that reduces the blurring in the cone-beam CT scan and helps the doctor to better locate the tumor. The investigators want to find out what effects, good and/or bad, that this has on the patient and the way we treat the cancer in the lungs. The second purpose of this study is to test how well this computer program tracks markers in the images, compared to using radio signals to follow the markers.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
17
Calypso transponders
Memorial Sloan Kettering Cancer Commack
Commack, New York, United States
Memorial Sloan Kettering Westchester
Harrison, New York, United States
Memorial Sloan Kettering Cancer Center
New York, New York, United States
successful Calypso transponder implantation
The RapidTrack algorithm will be considered successful if the centroid position of the Calypso transponders, as determined by the RapidTrack algorithm, are within 2 mm of the positions determined by the Calypso system 95% of the time.