The aim of the present study is to compare accuracy and precision of Cardiac Output (CO) by nine different pulse contour algorithms with transpulmonary thermodilution before and after cardiopulmonary bypass (CPB).
After induction of anesthesia, a central venous catheter and a transpulmonary thermodilution catheter (Pulsion Medical Systems, Munich, Germany) are introduced in the right internal jugular vein and in the femoral artery, respectively. The thermodilution catheter is connected to a PiCCO2 monitor (Software version 1.3.0.8). Thereafter, placement of an esophageal doppler probe is performed and the probe is connected to a unique Monitoring System (Deltex Medical, Chichester, UK) with nine different pulse contour algorithms. Cardiac Output by esophageal doppler is used to calibrate the nine different pulse contour algorithms. After induction of anesthesia and establishment of all monitoring devices, a passive leg raising maneouvre (PLR) is performed and hemodynamic variables including CO by transpulmonary thermodilution (COTPTD) and CO by nine different pulse contour algorithms (COX1-9) are recorded before, during and after PLR. Subsequently, measurements of COTPTD and COX1-9 are carried out every 10 minutes until the beginning of CPB. Fifteen minutes after weaning from cardiopulmonary bypass calibration of nine different pulse contour algorithms by esophageal doppler are performed again and measurements of CITPTD and CIPFX are restarted up to the end of the surgical Intervention.
Study Type
INTERVENTIONAL
Allocation
NA
Masking
NONE
Enrollment
30
Measurement of cardiac output by transpulmonary thermodilution, nine different pulse contour algorithms and esophageal doppler before, during and after a PLR-maneouvre and every 10 minutes before and after cardiopulmonary Bypass.
Berthold Bein
Hamburg, Free and Hanseatic City of Hamburg, Germany
Accuracy of cardiac output in l/min by nine different pulse contour algorithms compared with transpulmonary thermodilution
Time frame: After induction of anesthesia until the end of surgery
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.