The purpose of this study is to demonstrate the performance of the Harmony 1 Sensor in subjects age 2 - 75 years
This study is a multi-center, randomized, prospective single-sample correlational design without controls. A total of up to 310 previously-diagnosed type 1 or 2 diabetes subjects will be enrolled in order to have 160 subjects complete frequent sample testing at up to 17 centers. Subjects will wear Harmony 1 sensors used with the Guardian Mobile application.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Use of Harmony 1 Sensor over 10 days in arm and abdomen and used with the Guardian Mobile App in subjects aged 2-75 years who have had a diagnosis of type 1 or type 2 diabetes for at least one year.
Harmony 1 Sensor accuracy
Sensor accuracy will be evaluated compared to YSI plasma glucose values during frequent sample testing. A within 30% mean agreement rate (±22.5 mg/dL when Reference BG ≤ 75 mg/dL) between sensor and YSI will be evaluated against the null Hypothesis for varying glucose ranges.
Time frame: 10 days