The user performance evaluation shows whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system. In this study, user performance evaluation will be performed for Contour Plus (Bayer Healthcare Diabetes Care), Accu-Chek Active (Roche Diagnostics), Accu-Chek Performa (Roche Diagnostics) and OneTouch Select Simple (LifeScan) following ISO 15197:2013, clause 8. The study will be performed in 2 parts. Each part will comprise testing of 2 BGMS. For each BGMS, measurement procedures for user performance evaluation will be performed with 1 test meter and 1 reagent system lot by the study subjects. The same meter and an additional test meter will be used for double measurements performed by study personnel (with the same reagent system lot used by subjects).
Study Type
INTERVENTIONAL
Enrollment
150
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Ulm, Germany
Acceptance criteria defined by ISO 15197:2013 (see description)
95 % of the individual glucose measured values shall fall within ± 15 mg/dl (0.83 mmol/l) of the measured values of the comparison measurement procedure at glucose concentrations \< 100 mg/dl (5.55 mmol/l) and within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
Time frame: For each subject, the experimental phase has an expected duration of up to 3 hours