This study evaluates whether whole blood transferred through the new POLFA needle assembly meets supernatant hemoglobin acceptability standards.
The POLFA needle is part of the set's PPD (personal protection device) sampling port assembly that facilitates safer blood transfer between the collection set (i.e., the diversion pouch) and a Vacutainer® for the purpose of viral testing or other testing of the whole blood. The POLFA needle assembly was developed as a means to reduce costs while maintaining functionality, quality and safety. Since the overall functionality of the POLFA needle was designed to be equal to the currently approved Kawasumi needle, a change to the overall device safety and use profile is neither anticipated nor intended. This trial will serve to generate data sufficient to demonstrate that blood transferred through the POLFA needle results in plasma hemoglobin levels below 100mg/dL.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
OTHER
Masking
NONE
Enrollment
62
Each subject receives the same intervention and blood is drawn into both the INA and CNA. The difference is the order in which the CNA or INA is applied.
Each subject receives the same intervention and blood is drawn into both the INA and CNA. The difference is the order in which the CNA or INA is applied.
Haemonetics
Braintree, Massachusetts, United States
Hoxworth Blood Center
Cincinnati, Ohio, United States
American Red Cross Mid-Atlantic Region Blood Services
Norfolk, Virginia, United States
Plasma Supernatant Hemoglobin
To determine whether whole blood transferred to a Vacutainer® through the POLFA needle will have supernatant hemoglobin levels \<100 mg/dL within a 95% CI and 95% reliability. The average hemoglobin levels in the blood collected through each assembly is presented.