The purpose of this study is to conduct an oxygen saturation (SpO2) accuracy validation comparing prototype pulse oximeter oxygen saturation measurements directly to arterial blood saturation during non-motion conditions over the range of 60-100% SaO2 as measured by Reference CO-Oximetry. Additionally, the prototype pulse oximeter pulse rate measurements will be compared to 3-lead ECG heart rate measurements made via a multi-parameter monitor.
The description of this study is to conduct an oxygen saturation (SpO2) accuracy validation comparing prototype pulse oximeter oxygen saturation measurements directly to arterial blood saturation during non-motion conditions over the range of 60-100% SaO2 as measured by Reference CO-Oximetry. Additionally, the prototype pulse oximeter pulse rate measurements will be compared to 3-lead ECG heart rate measurements made via a multi-parameter monitor.
Study Type
OBSERVATIONAL
Clinimark
Louisville, Colorado, United States
The OxiCable pulse rate measurement performance is an ARMS of 3 or better over the full range 60-100% SaO2.
Time frame: 2.5 hours