The purpose of this study is to compare standard automated perimetry (SAP) using the Heidelberg Edge Perimeter (HEP) to the Octopus Visual Field (OVF) analyzer and determine test-retest reliability of both parameters in detecting glaucomatous visual field losses. This study will also investigate novel imaging parameters of the optic nerve head and new automated structure-function reports.
Each participant undergo the following tests: 1. Standard ophthalmic examination (usual care): Best-corrected visual acuity, biomicroscopy (look at front of eye), intraocular pressure (IOP) using Goldman applanation tonometry, central corneal thickness measurement and fundoscopy (look at back of eye). 2. Octopus Visual Field (OVF) Analyzer using G-top strategy recording mean deviation (MD) and pattern standard deviation (PSD). 3. Heidelberg Edge Perimeter (HEP) standard automated perimetry (SAP) using advanced staircase thresholding algorithm (ASTA). 4. Spectralis optical coherence tomography (OCT) using glaucoma module premium edition (GMPE) software to measure minimum rim width (MRW) and retinal nerve fiber layer (RNFL) thickness in 6 sectors (superior nasal, superior temporal, inferior nasal, inferior temporal, nasal and temporal) in the retina. Order of HEP SAP III and OVF, and order of the eye tested (right vs. left) if both eyes of patients meet inclusion criteria will be randomized. Patients with glaucoma will be randomly chosen and asked to return in 3 to 6 months for repeat HEP and OVF testing.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
115
HEP measures peripheral vision
OVF measures peripheral vision
OCT measures retinal nerve fiber layer thickness
Correlation Coefficient Between HEP and OVF Mean Deviation (MD)
Pearson's correlation coefficient between Heidelberg Edge Perimeter (HEP) and Octopus Visual Field (OVF) Mean Deviation (MD) for glaucoma patients and controls to determine if HEP can detect glaucoma as well as OVF. The closer the values of both parameters for both machines, the better comparable the two machines are to each other in detecting glaucoma.
Time frame: Baseline visit, 1 hour
Repeatability of Optical Coherence Tomography (OCT) Parameters
A subgroup of participants were randomly selected to return at 3 and 6 months for repeat testing. Not all eyes were included. Repeatability of optical coherence tomography (OCT) to consistently detect minimum rim width (MRW) global thickness will be assessed by Pearson's interclass correlation coefficients (ICC). A larger ICC indicates measurements have greater repeatability. Greater than 0.75 indicated excellent repeatability; 0.40 to 0.75 indicated fair to good, and less than 0.40 indicated poor reliability.
Time frame: Month 6 visit, 1 hour
Correlations Between Structure-function Automated Report and Clinical Impression
Heidelberg Edge Perimeter (HEP) Visual Field (VF) and spectral domain optical coherence tomography (SD OCT) printouts from only the glaucoma subgroup were assessed. Retinal Nerve Fiber Layer (RNFL) and Minimum Rim Width (MRW) from SD OCT and HEP VF automated reports were compared to the clinical interpretations by 3 glaucoma specialists.
Time frame: Baseline visit, 1 hour