The purpose of this study was to compare dynamic inspiratory flow rates achieved by a population of Chronic Obstructive Pulmonary Disease (COPD) patients through the Breezhaler®, Ellipta® and Handihaler® dry powder inhaler (DPI) devices. No active drug or placebo were administered to patients in this study.
The inspiratory flow rates achieved by the COPD patients, via the breath actuated devices, are one of the important factors that can influence the efficiency of the drug delivery by the inhalation route. The peak inspiratory flow (PIF) rates achieved by COPD patients through different breath actuated dry power inhalers may differ because of the differences in internal resistance of the different type of devices. The results of inspiratory airflow rate and pressure drop over time, from this study, provided an assessment of the range and variability of inhalation profile characteristics generated by COPD patients within the population selected for this study (e.g. demographics, gender, disease control) through the Breezhaler and the other marketed comparator DPIs. The inhalation profiles were obtained through an Inhalation Profile Recorder (IPR). The inhalation profile recorder was a data acquisition device which consists of a computer, an interface unit and a pressure transducer. This allowed it to measure the real time dynamic pressure drop at the mouthpiece of the inhaler during an inhalation maneuver. On the graphic user interface, plots of inspiratory pressure drop versus time and flow rate versus time are displayed in real time. The investigator or designated study personnel at site was responsible for typing in the Test Location, selecting the Inhaler Type (Breezhaler, Ellipta, Handihaler), typing in the Patient ID and Patient Age, selecting Male or Female, and recording Patient Comments made during the test session, as necessary. Each patient were required to generate three (03) inhalational profiles through each of the three (03) study devices: Breezhaler®, Ellipta® and Handihaler®.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
97
No active drug or placebo were administered to patients during this study. Each patient were required to inhale via all three (03) study devices (Breezhaler®, Ellipta® and Handihaler®). Each patient was required to produce at least three (03) inhalational profiles via each inhalational device, following the training and demonstration for correct inhalational procedure, for each device.
No active drug or placebo were administered to patients during this study. Each patient were required to inhale via all three (03) study devices (Breezhaler®, Ellipta® and Handihaler®). Each patient was required to produce at least three (03) inhalational profiles via each inhalational device, following the training and demonstration for correct inhalational procedure, for each device.
Novartis Investigative Site
CABA, Buenos Aires, Argentina
Novartis Investigative Site
Florida, Buenos Aires, Argentina
Novartis Investigative Site
La Plata, Buenos Aires, Argentina
Novartis Investigative Site
Mar del Plata, Buenos Aires, Argentina
Peak Inspiratory Flows Rates Summary by Inhalation Devices - PPS
The peak inspiratory flows (PIF) rates obtained from the inhalation flow profiles generated by the COPD patients through the three dry powder inhalation (DPI) devices(Breezhaler®, Ellipta® and Handihaler®) were measured and compared (without drug or placebo administration). Each patient were required to generate inhalation flow profiles through all three DPI devices in a randomized cross-over sequence. The inspiratory measurements were taken in each of these devices in the same visit.
Time frame: Visit 2 (Day 1)
Peak Inspiratory Flows Rates Summary by Inhalation Devices - FAS
The peak inspiratory flows (PIF) rates obtained from the inhalation flow profiles generated by the COPD patients through the three dry powder inhalation (DPI) devices(Breezhaler®, Ellipta® and Handihaler®) were measured and compared (without drug or placebo administration). Each patient were required to generate inhalation flow profiles through all three DPI devices in a randomized cross-over sequence. The inspiratory measurements were taken in each of these devices in the same visit. This FAS dataset includes PIF data from additional patients with corrected inhaler internal resistance values.
Time frame: Visit 2 (Day 1)
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
No active drug or placebo were administered to patients during this study. Each patient were required to inhale via all three (03) study devices (Breezhaler®, Ellipta® and Handihaler®). Each patient was required to produce at least three (03) inhalational profiles via each inhalational device, following the training and demonstration for correct inhalational procedure, for each device.
Novartis Investigative Site
Ciudad Autonoma de Bs As, Argentina