The study focuses on needle guidance in rheumatologic operations - glucocorticoid injections inside joint capsules specifically. The main purpose of the study is to assess and validate the clinical feasibility and performance of Bioimpedance Probe (BIP) Needles and assess the benefits, drawbacks and risks related to use of the BIP Needles.
Clinical device investigation
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
79
Injeq Bioimpedance Probe (BIP) Needle is an injection needle that has bioimpedance measurement capability. It consists of traditional needle cannulae and removable bioimpedance probe which enables the measurement of bioimpedance. The needle is connected to measurement device and tissue identifying algorithm. Bioimpedance is measured during the operation and the algorithm detects when the needle tip is in contact with synovial fluid.
Helsinki University Hospital, Clinic of Rheumathology
Helsinki, HUS, Finland
True Positive Detection of Synovial Fluid Marked in Case Report Form
Measurement device indicates with a sound and visual feedback when needle is in contact with synovial fluid. Physician verifies the location with ultrasound imaging, aspiration of synovial fluid and/or lack of resistance in glucocorticoid injection. Even if the device does not detect the synovial fluid, physician performs the injection when needed. Physician marks to the case report form whether the device provided detections during the puncture and were the detections true or false detections.
Time frame: During intra-articular injection