Nasal High Flow (NHF) and CPAP are established forms of respiratory support in neonates and often produce similar clinical outcomes. The aim of the study was to investigate airway pressure during NHF and CPAP in neonates.
Methods: A physiological study which measured pressure intra-nasally with micro (750 µm) pressure probes (Millar, USA) in 10 (5M / 5F) healthy newborns with mean gestation age 37.18 (SD 1.29) wk and mean birth weight of 2414 (SD 503.13) g. NHF was applied at 8 L/min (AIRVO 2) through smaller (OPT 314) and larger (OPT 316) Optiflow nasal cannula; CPAP 5 cmH20 was produced by a Bubble CPAP system and nasal mask interface at constant flow 8 L/min (Fisher \& Paykel Healthcare, NZ). Values are expressed in cmH20 as means and standard deviations (SD).
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
BASIC_SCIENCE
Masking
NONE
Enrollment
10
NHF 8 l/min by Airvo 2 by small cannula
NHF 8 l/min by Airvo 2 by large cannula
Nsal CPAP 5cm H2o by face mask
Neonatal Intensive Care Unit, Research Centre of Maternal and Child Health
Yerevan, Armenia
RCMCHP
Yerevan, Armenia
Upper airway pressure generated by NHF compared to CPAP
Pressure in upper airways
Time frame: Up to 1 hour