To evaluate the point accuracy of the FreeStyle Libre Flash Glucose Monitoring System when used at home by pregnant women with diabetes.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
83
Subjects will wear the FreeStyle Libre Flash Glucose Monitoring System masked for 14 days. During these 14 days subjects will be asked to perform 4 blood glucose tests (pre-meals and before bedtime) per day. At the same time as the blood glucose test a sensor scan is performed.
VIVIT-Institut am Akad. Lehrkrankenhaus Feldkirch, Innere Medizin und Kardiologie
Feldkirch, Austria
Medizinische Universität Graz
Graz, Austria
Medizinische Universität Wien
Vienna, Austria
Point Accuracy Determined as % Within Consensus Error Grid Zone A
Point accuracy of Sensor based glucose values versus fingerstick blood glucose determined as % within Consensus Error Grid zone A. Zone A is defined as the zone of "clinical accurate measurements with no effect on clinical action."
Time frame: 14 days