To verify that platelets collected on the Trima Accel system with Version 7.0 software enhancements and stored in platelet additive solution (PAS) meet the FDA requirements for leukoreduction (\< 5.0 × 10\^6 residual white blood cells \[WBC\] per transfusable unit).
The study is designed to evaluate changes to the Trima Accel software to ensure the modified software meets the FDA acceptance criteria for leukoreduction and platelet yield. The operating range for the system flow rates, anticoagulant ratios, storage conditions, and centrifugal forces are the same as the currently cleared Trima Accel system. There are no changes to the environment or storage conditions for platelets, therefore, no in vitro or in vivo platelet quality data will be collected. This is a prospective, open-label, multi-center controlled study to evaluate the leukoreduction of platelets stored in platelet additive solution (PAS) collected on the Trima Accel system Version 7.0 software enhancement. Up to 450 healthy adult subjects will be enrolled in this study to ensure 93 single, 93 double and 93 triple platelet product evaluable data points. Evaluable is defined as a completed platelet product that does not meet any of the protocol analysis exclusion criteria. Plateletpheresis will be per site standard practice and per applicable FDA guidelines such as the FDA Collection of Platelets by Automated Methods, December 2007 Guidance for Industry and Pre-storage Leukocyte Reduction of Whole Blood and Blood Components Intended for Transfusion, September 2012. Study participation will be up to 30 days and will consist of 1 to 2 visits. Screening may be done within 30 days before the apheresis procedure or combined as a single visit, which includes screening and the apheresis procedure all in 1 day. The entire study should be completed in approximately 16 months.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
OTHER
Masking
NONE
Enrollment
365
Platelet Apheresis Procedure
San Diego Blood Bank
San Diego, California, United States
Bonfils Blood Center
Denver, Colorado, United States
Indiana Blood Center
Indianapolis, Indiana, United States
Memorial Blood Centers
Minneapolis, Minnesota, United States
Number of Participants With Platelet Units Containing an Acceptable Residual WBC Level
The number of participants with platelet units containing an acceptable residual WBC level. Acceptable residual WBC counts are: singles = residual WBC level \< 5.0 × 10\^6; doubles = residual WBC level \< 8.0 × 10\^6 or \< 5.0 × 10\^6 for each transfusable unit; and triples = residual WBC level \< 12.0 × 10\^6 or \< 5.0 × 10\^6 for each transfusable unit.
Time frame: Within 48 hours of end of donation
Number of Participants With Platelet Units Containing an Acceptable Platelet Yield
The number of participants with platelet units containing an acceptable platelet yield. Acceptable platelet yield for single, double and triple platelet products are: platelet yield ≥ 3.0 × 10\^11 for singles, platelet yield ≥ 6.2 × 10\^11 for doubles, and platelet yield ≥ 9.3 × 10\^11 for triples.
Time frame: Within 48 hours of end of donation