The study is intended to demonstrate substantial equivalence of visualization using BLI to the predicate device, FICE, so as to support the addition of BLI as an adjunct to white light imaging in the company's video processor.
This clinical study will consist of acquisition of images, followed by a reader study.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
20
All subjects will have 1 endoscopy procedure, in which they will have a white light examination followed by FICE and BLI imaging on a maximum of two (2) anatomical sites per subject.
Mayo Clinic AZ
Scottsdale, Arizona, United States
Image Quality Scores Using Likert Scale
Likert score ratings of image quality for Blue Light Imaging (BLI), Blue Light Imaging-bright (BLI-brt) and Flexible spectral-Imaging Color Enhancement (FICE) images. Each BLI and BLI-Brt setting was compared to FICE, to determine which settings were equivalent to FICE. Likert Score Scale: 1. Not seen; 2. poor but usable, characteristic features are detectable but details are not fully reproduced; features just visible; 3. good; allows an adequate assessment, details of anatomical structures are visible but not necessarily clearly defined; details emerging; 4. very good; allows an excellent assessment, anatomical details are clearly defined; details clear This was a blinded and randomized Reader assessment of all BLI and FICE images.
Time frame: Two to three months