A comparative study was conducted and the patient underwent a conventional urodynamic study. In order to successfully determine if the Air-Charged (AC) and Water-Perfused (WP) measurements are equivalent, the two sources of intravesical pressure (Pves) and abdominal pressure (Pabd) were collected concurrently at various fill volumes for the bladder.
The objective of this study was to compare the use of water-filled and air-charged catheters in determining equivalency between the two technologies during cystometric assessment. A total of 25 patients (9M/16F) were recruited. All patients underwent cough and Valsalva manoeuvre pressure tests to measure vesicle pressure (Pves) and abdominal pressure (Pabd). A single dual-lumen catheter (T-DOC 7Fr Air-Charged® catheter) was used to record air and water pressures simultaneously. The primary outcome was to determine if the maximum pressures during Valsalva manoeuvres, as measured with a single dual-lumen water-perfused and air-charged catheter, are equivalent when the bladder is filled to 200 cc during a urodynamic evaluation. Exploratory endpoints included the following: 1. To determine if maximum pressures for cough, as measured with a single dual-lumen water-perfused and air-charged catheter, are equivalent when the bladder is filled to 100, 200 and Maximum Bladder Capacity (MBC) cc during a urodynamic evaluation. 2. To determine if maximum pressures for Valsalva manoeuvres, as measured with a single dual-lumen water-perfused and air-charged catheter, are equivalent when the bladder is filled to 100 and MBC cc during a urodynamic evaluation (as well as 200 cc which is the primary objective). 3. To determine if the maximum voiding pressure, as measured with a single dual-lumen water-perfused and air-charged catheter, are equivalent. 4. To determine if the clinical impressions of the urodynamic study are equivalent for the water-perfused and air-charged catheters. 5. To determine if the compliance of the bladder is equivalent when measured by air-charged catheters as compared to water-perfused catheters.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
25
Urodynamic study utilizing a single catheter technique to measure WP \& AC measurements
University Clinic Bonn
Bonn, Germany
Maximum Vesical Pressure During Valsalva Manoeuvres
Maximum vesical pressure during Valsalva manoeuvres was measured with an air-filled catheter and the water pressure was measured from the fill port. Air-filled catheter and water pressure measurements were compared.
Time frame: Measured during a single urodynamic evaluation
Maximum Abdominal Pressure During Valsalva Manoeuvres
Maximum abdominal pressure during Valsalva manoeuvres was measured with an air-filled catheter and the water pressure was measured from the fill port. Air-filled catheter and water pressure measurements were compared.
Time frame: Measured during a single urodynamic evaluation
Maximum Detrusor Pressure at Valsalva Manoeuvres
Maximum detrusor pressure during Valsalva manoeuvres was measured with an air-filled catheter and the water pressure was measured from the fill port. Air-filled catheter and water pressure measurements were compared.
Time frame: Measured during a single urodynamic evaluation
Maximum Vesical Pressure During Coughs
Maximum vesical pressure during coughs was measured with an air-filled catheter and the water pressure was measured from the fill port. Air-filled catheter and water pressure measurements were compared.
Time frame: Measured during a single urodynamic evaluation
Maximum Abdominal Pressure During Coughs
Maximum abdominal pressure during coughs was measured with an air-filled catheter and the water pressure was measured from the fill port. Air-filled catheter and water pressure measurements were compared.
Time frame: Measured during a single urodynamic evaluation
Maximum Detrusor Pressure During Coughs
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Maximum detrusor pressure during coughs was measured with an air-filled catheter and the water pressure was measured from the fill port. Air-filled catheter and water pressure measurements were compared.
Time frame: Measured during a single urodynamic evaluation