Background During ventilatory assistance, optimization of settings is critical to allow a personalized support and avoid over- or under-assistance. But little data are available in clinical practice to guide the adjustment of the support. In adults, esophageal pressure (PES) has been shown to be a reliable surrogate of pleural pressure (PPL) and clinical studies suggest that PES may be useful to guide the management of mechanical ventilation. In children, the PES measurement could have similar potential benefits, but beforehand the reliability of PES to estimate PPL needs to be assessed. Objective The primary objective of this study is to validate the reliability of PES directly monitored using a miniature catheter tip pressure transducer (Gaeltec® system) to estimate PPL, when compared to a gold standard, i.e the direct PPL measurement in situ. Method This is a prospective single center study. Children \<18 years old, hospitalized in the pediatric intensive care unit, requiring invasive ventilation and with at least one chest tube will be included. Protocol A pressure transducer will be connected to the existing chest-tube and PES (measured by Gaeltec® and feeding tube), PPL, PAW, respiratory volume and flow will be simultaneously recorded. Expected results We expect that the PES-based methods will provide an accurate estimation of PPL. Once this tool validated, PES could be helpful to optimize mechanical ventilation in children, and further interventional trials would be warranted to evaluate if its use could allow a reduction of the ventilation support duration.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
BASIC_SCIENCE
Masking
NONE
Enrollment
12
Esophageal pressure will be monitored by a dedicated catheter and pleural pressure will be measured into the chest tube already in place
St. Justine's Hospital
Montreal, Quebec, Canada
End-expiratory esophageal
End-expiratory PES and PL, calculated as the mean value of 10 consecutive breaths, during a stable period with no intervention or artifacts;
Time frame: Up to 2 hours from the beginning of the study
Pleural pressure
PPL, directly measured in the existing chest tubes, and calculated as the mean value of 10 consecutive breaths, during a stable period with no intervention or artifacts.
Time frame: Up to 2 hours from the beginning of the study
transpulmonary pressures
Time frame: Up to 2 hours from the beginning of the study
The Elastance-derived end-inspiratory transpulmonary pressure
The Elastance-derived end-inspiratory transpulmonary pressure, calculated as: PAW at end-inspiration x EL/ERS (EI and ERS being the respective elastances of the lung and the respiratory system, estimated on the esophageal and airway Pressure-Volume curves);
Time frame: Up to 2 hours from the beginning of the study
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.