The purpose of this study is to establish the safety and feasibility of the Edwards SAPIEN 3 valve in subjects with mitral annular calcification (MAC) associated with mitral stenosis (MS) and/or mitral regurgitation who are at high-risk for mitral valve surgery or deemed inoperable due to the extent of calcification.
Technical Success: Alive, with * Successful access, delivery and retrieval of the device delivery system, and * Deployment and correct positioning (including repositioning/recapture if needed) of the single intended device, and * No need for additional unplanned or emergency surgery or re-intervention related to the device or access procedure Device Success: Alive and stroke free, with * Original intended device in place, and * No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/Operating Room(OR)), and * Intended performance of the device: * Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and * Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = Mitral Valve Area (MVA) \< 1.5cm2 and Mitral Valve (MV) gradient \> 5mmHg, Insufficiency = Mitral Regurgitation (MR) \>1+), and * Absence of para-device complications (e.g., Paravalvular Leak (PVL) \> mild, need for a Permanent Pacemaker (PPM), erosion, Annular rupture or Aortic Valve (AV) Groove disruption, Left Ventricular Outflow Tract (LVOT) gradient increase \> 10mmHg) Procedural Success: Device success, and * No device or procedure related Serious Adverse Events (SAEs) (Life threatening bleed; major vascular or cardiac structural complications requiring unplanned reintervention or surgery; stage 2 or 3 Acute Kidney Injury (includes new dialysis); Myocardial Infarction (MI) or need for percutaneous coronary intervention (PCI)/coronary artery bypass graft (CABG); severe heart failure (HF) or hypotension requiring intravenous (IV) inotrope, ultrafiltration or mechanical circulatory support; prolonged intubation ( \> 48 hours) 6.2 Secondary objective(s) Device Success (at 6 months and 1 year) * Subject success 1 year Device Success: Alive and stroke free, with * Original intended device in place, and * No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/Operating Room (OR)), and * Intended performance of the device: * Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and * Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = MVA \< 1.5cm2 and MV gradient \> 5mmHg, Insufficiency = MR \>1+), and * Absence of para-device complications (e.g., PVL \> mild, need for a PPM, erosion, Annular rupture or AV Groove disruption, LVOT gradient increase \> 10mmHg)
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
25
subjects with surgical MVR with Sapien3
St. Vincent Heart Center
Indianapolis, Indiana, United States
MHRI Maryland
Baltimore, Maryland, United States
Lankenau Institute for Medical Research
Wynnewood, Pennsylvania, United States
Baylor College of Medicine
Houston, Texas, United States
Procedural Success
Procedural Success is defined as: 1. Device success, and 2. No device or procedure related serious adverse events (SAEs) (Life threatening bleed; major vascular or cardiac structural complications requiring unplanned reintervention or surgery; stage 2 or 3 acute kidney injury (AKI) (includes new dialysis); myocardial infarction (MI) or need for percutaneous coronary intervention (PCI)/coronary artery bypass graft (CABG); severe heart failure (HF) or hypotension requiring intravenous (IV) inotrope, ultrafiltration or mechanical circulatory support; prolonged intubation ( \> 48 hours)
Time frame: 30 days
Technical Success
Technical Success is defined as: 1. Alive, with 2. Successful access, delivery and retrieval of the device delivery system, and 3. Deployment and correct positioning (including repositioning/recapture if needed) of the single intended device, and 4. No need for additional unplanned or emergency surgery or re-intervention related to the device or access procedure
Time frame: Procedure Stop Time
Device Success
Device Success is defined as: 1. Alive and stroke free, with 2. Original intended device in place, and 3. No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/operating room (OR)), and 4. Intended performance of the device: 1. Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and 2. Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = mitral valve area (MVA) \< 1.5cm2 and mitral valve (MV) gradient \> 5mmHg, Insufficiency = mitral regurgitation (MR) \>1+), and 3. Absence of para-device complications (e.g., paravalvular leak (PVL) \> mild, need for a permanent pacemaker (PPM), erosion, Annular rupture or aortic valve (AV) Groove disruption, left ventricular outflow tract (LVOT) gradient increase \> 10mmHg)
Time frame: 30 days
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
The Heart Hospital Baylor Plano
Plano, Texas, United States
University of Virginia
Charlottesville, Virginia, United States
Device Success
Device Success is defined as: 1. Alive and stroke free, with 2. Original intended device in place, and 3. No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/operating room (OR)), and 4. Intended performance of the device: 1. Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and 2. Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = MVA \< 1.5cm2 and MV gradient \> 5mmHg, Insufficiency = MR \>1+), and 3. Absence of para-device complications (e.g., PVL \> mild, need for a PPM, erosion, Annular rupture or AV Groove disruption, LVOT gradient increase \> 10mmHg)
Time frame: 6 months
Device Success
Device Success is defined as: 1. Alive and stroke free, with 2. Original intended device in place, and 3. No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the cath lab/OR), and 4. Intended performance of the device: 1. Structural performance: No migration, embolization, detachment, fracture, hemolysis, thrombosis (including reduced leaflet motion) or endocarditis, etc, and 2. Hemodynamic performance: Maintenance of relief of stenosis or insufficiency without producing the opposite (Stenosis = MVA \< 1.5cm2 and MV gradient \> 5mmHg, Insufficiency = MR \>1+), and 3. Absence of para-device complications (e.g., PVL \> mild, need for a PPM, erosion, Annular rupture or AV Groove disruption, LVOT gradient increase \> 10mmHg)
Time frame: 1 year
Subject Success
Subject Success is defined as: All of the following must be present: I. Device success (either optimal or acceptable), and II. Patient returned to the pre-procedural setting: and III. No rehospitalizations or reinterventions for the underlying condition (e.g., mitral regurgitation, heart failure); and IV. Improvement from baseline in symptoms (e.g., NYHA improvement by \> or = 1 functional class); and V. Improvement from baseline in functional status (e.g., 6-min walk test improvement by \> or = 50 m); and VI. Improvement from baseline in quality-of-life (e.g., Kansas City Cardiomyopathy Questionnaire improvement by \> or = 10)
Time frame: 1 year