The position of the RV lead may have an important role in the effectiveness of the Programmable Hypertension Control (PHC) therapy by the Moderato pacemaker eliciting a lag from which RV signals (QRS) are identified. The study attempts to measure RV Pressure and RV Volume with PHC in two RV lead locations.
Subjects found eligible for participation in the CS-03 study, who are indicated for a de novo pulse generator implant at the center participating in this sub-study and who are not 100% dependent on RV pacing are eligible for participation in this sub-study. Patients receiving a pacemaker pulse generator exchange or who are 100% pacemaker dependent may not participate. Eligible patients will be provided with an Informed Consent specific to this sub-study and if patients choose to participate, patients will be asked to sign the Informed Consent and will then undergo the procedures described below. Patients will be prepared and draped for the Moderato implant procedure as detailed in the main CS-03 study protocol. Subjects will be instrumented prior to the Moderato Pacemaker implant procedure. The position of the RV lead will be tested at two sites known to elicit shortest QRS duration (The RV apical septum and the His region). A series of three test periods with the Moderato System programmed to deliver predetermined PHC settings will follow. Following the completion of these measurements, the conductance catheter will be withdrawn and the Moderato System implant will resume as described in the CS-03 protocol.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
BASIC_SCIENCE
Masking
NONE
Enrollment
5
A conductance catheter will be used to obtain PV loops prior to a Moderato pacemaker implant procedure.
Na Homolce Hospital
Prague, Czechia
Change in the LV volumes (ml)
LV volume will be assessed in two RV electrode lead locations with BackBeat PHC ON and OFF
Time frame: up to 2 hr
Change in the LV pressures (mmHg)
LV pressure will be assessed in two RV electrode lead locations with BackBeat PHC ON and OFF
Time frame: up to 2 hr
Change in the pressure-volume ratio
LV pressure will be assessed in two RV electrode lead locations with BackBeat PHC ON and OFF
Time frame: up to 2 hr