E-Celsius device used for measuring the gastrointestinal temperature is composed of an ingestible capsule measuring the temperature as well as a transportable module receiving data from the capsule. This device allows to measure the body temperature (± 0.2 ° C) during transit of the pill in the digestive tract to a maximum sampling frequency of 30 seconds. The main risks related to such devices are related to biocompatibility (in materials used for the capsule shell, or level of contamination of the gastrointestinal environment due to a defect in the plastic envelope the pill or a lack of hygiene, previously tested in clinical trials; see section 10.2). It will compare conventional methods of estimating the central temperature continuously with gastrointestinal data obtained from the e-Celsius device. The main objective of the research is to validate the extent of gastrointestinal temperature obtained from the device under test consists of an ingestible electronic capsule and a data reception monitor. Secondary objectives will be pursued: * Assess the effects of ingestion of cold drinks on temperature readings taken via e-device Celsius, esophageal probe and a rectal probe. * Explore the thermal homogeneity of the digestive tract. * Evaluate the internal memory of the capsule and its data forwarding capacity. This protocol will also measure the reliability of the measurement system while comparing it to the standards used by health staff today.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
BASIC_SCIENCE
Masking
NONE
Enrollment
22
comparison of temperature between e-celsius and thermometer, rectal thermometer, esophageal thermometer, vital sense capsule thermometer
Caen University Hospital
Caen, France
body temperature
comparisons between device e-Celsius, esophageal probe and a rectal probe
Time frame: from baseline to 36 hours every minute
body temperature during ingestion of cold drinks
Assess the effects of ingestion of cold drinks on temperature readings taken via e-device Celsius, esophageal probe and a rectal probe
Time frame: baseline
temperature of the digestive tract
Explore the thermal homogeneity of the digestive tract
Time frame: from baseline to 36 hours every minute
number of lost data
The capacity of the for each measurements method to store temperature data in the internal memory and send it to the receiver when the it is in the receiving area.
Time frame: from baseline to 36 hours every minute
reliability of measurements
reliability of the measurement system while comparing it to the standards used by health staff today
Time frame: from baseline to 36 hours every minute
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.