The purpose of this study is to evaluate the accuracy of the Dario Blood Glucose Monitoring System as compared to a reference method (Yellow Springs Instrument 2300 STAT Plus) and to evaluate the usability of the device when used by laypersons.
The objectives for this study are to evaluate the blood glucose level results obtained from the subject's fingertip using the Dario Blood Glucose Monitoring System paired with a reference device from one of three Android smart mobile device families as compared to the results from the Yellow Springs Instrument 2300 STAT Plus. The usability of the Dario Blood Glucose Monitoring System when paired with a reference device from one of three Android smart mobile device families and used by laypersons is also evaluated.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
102
Accuracy with usability assessment
Barbara Davis Center
Aurora, Colorado, United States
Accuracy of Glucose Concentrations (%)
Accuracy acceptability is defined as follows: Ninety-five percent (95%) of the individual glucose results shall fall within +/-15 mg/dL of the results of the Dario's measurement at glucose concentrations \< 75 mg/dL and within +/- 20% at glucose concentrations greater than or equal to 75 mg/dL. The accuracy of the Dario Blood Glucose Monitoring System when used with Android operating system smart mobile devices was assessed against the Yellow Springs Instrument 2300 STAT Plus.
Time frame: 1 Day
Average Rank of the Indication of "How Would Rate Your Success in Operating Dario BGMS"
Using a scale of 1-5 where 1-completely fails, 2-unsuccessful, 3-Partially successful, 4-successful, 5-very successful. Average rank of 3.5 ('average'/'successful') or above in question #2 - How would you rate your success in operating Dario™ BGMS?
Time frame: 1 Day
Average Rank of the Indication of "How Easy Was it to Operate the Dario™ BGMS on the First Time?"
Using a scale of 1-5 where 1-very hard, 2-hard, 3-average, 4-easy, 5-very easy. An average rank of 3 ('average') or above in question #3 - How easy was it to operate the Dario™ BGMS on the first time?
Time frame: 1 Day
Percent of Participants Responding 'YES' on the Question "Did You Successfully Obtain Measurement Result Using the Dario™ BGMS?"
The acceptance criteria for this question is: at least ninety percent (90%) need to answer 'Yes' on question #1 - Did you successfully obtain measurement result using the Dario™ BGMS? 100% of subjects (102 of 102) answered "Yes" to this question.
Time frame: 1 Day
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.