Patients who are intubated and mechanically ventilated for acute respiratory failure in the Intensive Care Unit (ICU) are at some point eligible for weaning. The common way to wean them from mechanical ventilation is to screen criteria for feasibility and, if present, to test feasibility by performing spontaneous breathing trial. This latter can be done either by setting a low pressure support level (expected to compensate the airflow resistance due to endotracheal tube) or by allowing the patient to breathe spontaneously through the tube without any support from the ventilator. Combination of low pressure assistance strategy (7 cm H2O) and positive expiratory pressure (PEP) of 4 cm H2O is the strategy used in our unit. Such a low pressure support level should actually result in a real assistance and, hence this is not the real spontaneous breathing capacity that is tested. Some ICU ventilators offer the option of compensating for the airflow resistance due to endotracheal tube, automatic tube compensation (ATC). Therefore, investigators aimed at comparing in patients ready to wean the usual procedure in our ICU and the ATC mode. In the ATC arm, the patients are breathing spontaneously through the endotracheal tube and are connected to the ventilator set at inspiratory pressure support of 0 cm H2O, PEP 4 cm H2O and ATC on. Two parallel arms depending on the order of allocation of each mode: pressure support 7 cm H2O + PEP 4 cm H2O then ATC or the opposite. The primary endpoint is the power of the work of breathing. The hypothesis is that the power of the work of breathing is greater in ATC than in the usual procedure, and hence this latter is a real ventilator support.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
20
spontaneous breathing through endotracheal tube with no ventilator support except for ATC, the patient is connected to the ventilator with pressure support 0 cm H2O and PEP 4 cm H2O during 30 minutes then 30 minutes of basal pressure support then 30 minutes with pressure support 7 cm H2O PEP 4 cm H2O
30 minutes with pressure support 7 cm H2O PEP 4 cm H2O then 30 minutes of basal pressure support then 30 minutes with spontaneous breathing through endotracheal tube with no ventilator support except for ATC, the patient is connected to the ventilator with pressure support 0 cm H2O and PEP 4 cm H2O
Hôpital de la Croix Rousse
Lyon, France
the power of work of breathing
the work of breathing is measured from the area subtended by the esophageal pressure - lung volume relationship. The power of work of breathing is obtained by multiplying work of breathing by the respiratory frequency and expressed in Joules/min
Time frame: 30 minutes after onset the ventilator strategy
Respiratory rate
the respiratory rate is measured from the airflow tracing
Time frame: 30 minutes
tidal volume
integration of airflow over time during expiration
Time frame: 30 minutes
intrinsic PEP
deflection in esophageal pressure up to the first zero flow
Time frame: 30 minutes
occlusion pressure at 100 ms
measured from the airway pressure at the first 100 ms after airway occlusion
Time frame: 30 minutes
work of breathing per liter
work of breathing normalized for 1 liter tidal volume
Time frame: 30 minutes
Distribution of pulmonary ventilation
distribution of pulmonary ventilation will be measured by the Pulmovista system
Time frame: 30 minutes
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.