The goal of the proposed study is to test the capability of the CareTaker monitor to extract the respiration signature from its acquired signal, and to determine a respiration rate of breaths/minute that meets or exceeds the performance levels of its predicate device.
The goal of the proposed study is to test the capability of the CareTaker monitor to extract the respiration signature from its acquired signal, and to determine a respiration rate of breaths/minute that meets or exceeds the performance levels of its predicate device. The detection approach of RSPR (respiratory rate) that the CareTaker technology takes is based on peak detection and pulse analysis of arterial pressure pulse signals, as well as spectral analysis of the raw CareTaker sensor signal. The study was performed under the supervision of the Chesapeake IRB and involved 21 subjects \>18 y who were monitored during spontaneous and metronome-paced breathing.
Study Type
OBSERVATIONAL
Enrollment
21
Equivalence of CareTaker's respiration rate detection capability to that of predicate device
Respiration rate detection with accuracy of better than 3 breaths per minute
Comparison with performance of predicate device (accuracy within 3 breaths per minute).
Time frame: 15 minutes of matched respiration rates