Prospective, open-labeled, randomized, controlled, multicenter study at two clinical investigative sites in Tucson, Arizona. The subjects will undergo intraocular pressure measurement with the CATS tonometer prism and the Goldmann applanation tonometer prism (GAT). Of the 200 eyes, at least 10 highly astigmatic eyes (\>3 D of corneal astigmatism) each in the low, medium and high IOP ranges shall have the same clinical testing and analysis of results, with the analysis kept separate from that of the main group.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
86
demonstrate that not more than 5% of the paired differences between the reference tonometer (CATS prism) reading and the test tonometer (Goldmann prism) reading for each pressure range
demonstrate that not more than 5% of the paired differences between the reference tonometer (CATS prism) reading and the test tonometer (Goldmann prism) reading for each pressure range
Arizona Eye Consultants
Tucson, Arizona, United States
All subjects will have CATS tonometer prism intraocular pressure readings compared to Goldmann tonometer prism readings in establishing effectiveness.
Time frame: From date of randomization until 24 hours