The purpose of this study was to conduct a side by side comparison of the accuracy of respiratory rate (RR) between our investigational Rainbow Acoustic Monitoring (RAM) small sensor and the FDA-cleared RAM Revision D sensor in healthy volunteers under controlled conditions.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
29
Noninvasive acoustic monitoring device
Masimo Clinical Lab
Irvine, California, United States
Performance comparison of RAM sensors by ARMS calculation
Performance comparison was determined by comparing the noninvasive respiration rate measurement of the RAM sensors and calculating the arithmetic root mean square (Arms) error value.
Time frame: 50 minutes