Evaluation of three left ventricular pacing polarities effectiveness to narrow phrenic nerve stimulation.
The primary endpoint of the study aims at demonstrating that the three left ventricular polarities in the Cardiac Resynchronisation Therapy-Defibrillator could bypass phrenic nerve stimulation at least in 90% of implants. Success was defined as an absence of phrenic nerve stimulation (threshold \> 7V) or in case of phrenic nerve stimulation occurrence (threshold \< 7V), as the resolution of the phrenic nerve stimulation by reprogramming one of the 3 left ventricular pacing polarities available in the device : left ventricular tip- left ventricular ring (bipolar), left ventricular tip- right ventricular ring (pseudo-bipolar) and left ventricular ring-right ventricular coil. Any occurrence of phrenic nerve stimulation without resolution by reprogramming was considered as a failure.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
PREVENTION
Masking
NONE
Enrollment
90
Patients were implanted with a Cardiac Resynchronization Therapy -Defibrillator (model PARADYM RF SONR CRT-D 9770 or PARADYM RF CRT-D 9750, Sorin) having 3 left ventricular polarities. The choice of the right atrial, right ventricular and bipolar left ventricular leads was left to investigators' discretion.
CH Annecy
Annecy, France
Hopital Henri Mondor
Créteil, France
CH LENS
Lens, France
CH de Bretagne Sud
Lorient, France
Phrenic nerve stimulation measurement
The presence of phrenic nerve stimulation at 10 Volt, using a pacing system analyzer
Time frame: The presence of phrenic nerve stimulation will be assessed at implant.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Clinique Clairval
Marseille, France
CHP Beauregard
Marseille, France
Clinique Montpellier
Montpellier, France
Polyclinique de Gentilly
Nancy, France
CHU Groupe Hospitalo-universitaire Caremeau
Nîmes, France
Hopital de La Source
Orléans, France
...and 2 more locations