Purpose: Characterize the depositional attributes of a specialized trans-nasal aerosol delivery device in healthy people using clinical measurements of mucociliary clearance. Procedures (methods): 14% hypertonic saline (HS) aerosol containing the radio tracer Tc99m-sulfur colloid will be delivered via the tPAD and compared to the well characterized PARI LC Star nebulizer (7% HS) using planar gamma scintigraphy (i.e. particle clearance). The fraction of sulfur colloid particles that clear over the ensuing 24 hours with each device will indicate the fraction deposited in conducting airways.
Purpose: This is an early device characterization study to define the fraction of delivered aerosol generated by a novel nebulizer (tPAD) that deposits in conducting airways. Design: Open label, cross-over design, using a standard oral nebulizer (PARI LC Star) as a comparator. The order of visits will be randomized. Methods: Gamma scintigraphy will be performed following use of each device on separate study visits. Primary Outcome: Fraction of inhaled particles that clears over 24 hours.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DEVICE_FEASIBILITY
Masking
NONE
Enrollment
12
A novel, transnasal aerosol generator
Standard oral nebulizer
University of North Carolina at Chapel Hill
Chapel Hill, North Carolina, United States
24-hour clearance fraction
The percentage of initially deposited aerosol particles that clear over 24 hours will be measured with gamma scintigraphy after 24 hours
Time frame: 24 hours
Central:Peripheral Deposition Ratio
The ratio of labelled particles initially deposited in the central and peripheral lung zones will be assessed to characterize aerosol targeting of the lung
Time frame: 0-4 minutes after deposition