This study will use contrast echocardiography to assess the accuracy of a new non-invasive imaging method for subharmonic aided pressure estimation (SHAPE) compared to simultaneously acquired intra-cardiac pressures measured invasively during cardiac catheterization. This study is designed to verify that contrast echocardiography using the SHAPE method, already proven in a canine model and tested in a human pilot study can be used as a surrogate for cardiac catheterization with sufficient accuracy to allow clinical applicability in humans.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
80
Whenever a patient undergoes cardiac catheterization, which routinely includes intracardiac pressure monitoring using a pressure catheter, as part of their standard clinical care and agrees to participate in the study, we will acquire research data. With Sonazoid infusion, SHAPE algorithm will be initiated to determine optimum acoustic pressure. Then, at the optimum acoustic pressure, SHAPE specific data will be acquired from the cardiac chambers and aorta synchronously with the pressures recorded by the catheter (as a part of the patient's standard of care). After acquiring the ultrasound imaging data, the remainder of the heart catheterization will be completed by the attending cardiologist according to the patients' standard of care.
Thomas Jefferson University and Thomas Jefferson University Hospital
Philadelphia, Pennsylvania, United States
Agreement Between SHAPE and Pressure Catheter Measurements
The Pearson correlation coefficient was computed between the simultaneously acquired subharmonic signal and pressure catheter data (i.e., waveforms) for each participant. The Pearson correlation coefficient was used for comparing waveforms given the underlying data and pressure values used for comparisons followed a normal distribution. Values reported will be the mean Pearson correlation coefficient values between the subharmonic waveforms and pressure catheter waveforms obtained.
Time frame: up to 1 day
Error Between SHAPE and Pressure Catheter Measurements
Based on the distribution of the differences in cardiac pressure values obtained using the SHAPE technique and the clinical pressure catheter, either a two-tailed paired t test (normal distribution) or Wilcoxon signed rank test (nonnormal distribution) was used to compare the SHAPE technique to the reference standard. Bonferroni corrections were used for multiple comparisons
Time frame: up to 1 day