The present study aims to evaluate patient satisfaction and audiological performance with their current cochlear implant speech processor, the Neuro 1, and to compare it with the satisfaction and performance obtained with the new speech processor, the Neuro 2.
Nowadays, cochlear implants (CI) are successfully used to rehabilitate severe to profound sensorineural hearing loss. Oticon Medical recently developed a new cochlear implant system, including a new implantable internal part (Neuro Zti) and a new speech processor (Neuro 1). This new speech processor automatically detects and adapts to the sound environment, increasing the ease of listening in all situations. A second version of this speech processor has very recently been developed, called the Neuro 2. The Neuro 1 and 2 share the same signal processing characteristics and the "sound" delivered to the patient is similar. However, the Neuro 2 offers better ergonomics (smaller size and weight, rechargeable batteries…). Some improvements available on the Neuro 2 may be particularly interesting for children. The aim of the present study is twofold: 1. Compare the satisfaction of adult and pediatric users for the speech processor Neuro 1 with their satisfaction for the Neuro 2, and 2. Show that audiological outcomes are at least as good for the Neuro 2 than for the Neuro 1. The participants come at the hospital for three visits. The Neuro 1 is evaluated during the first visit (V1). Evaluation of the Neuro 2 takes place during the second (V2, 15 days after V1) and the third (V3, 3 months after V2).
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
98
Participants have to complete a satisfaction questionnaire on the Neuro 1 or 2 processor.
The speech material corresponds to the lists of Lafon designed for cochlear implant patients. Each lists contains 17 monosyllabic words. Speech comprehension is evaluated by the percentage of correct words and the percentage of correct phonemes. Speech signals are presented at 65 dB SPL. 2 lists of 17 words are presented. At visit 1, the participants wear the sound processor Neuro 1. At visits 2 and 3, the participants wear the Neuro 2.
The speech material corresponds to the lists of Lafon designed for cochlear implant patients. Each lists contains 17 monosyllabic words. Speech comprehension is evaluated by the percentage of correct words and the percentage of correct phonemes. Speech signals are presented at 65 dB SPL. 2 lists of 17 words are presented. The noise corresponds to a cocktail party presented at +10 dB SNR (Speech on Noise Ratio). At visit 1, the participants wear the sound processor Neuro 1. At visits 2 and 3, the participants wear the Neuro 2.
Hôpital Pellegrin
Bordeaux, France
Hôpital Roger Salengro
Lille, France
Hôpital Edouard Herriot
Lyon, France
CHRU de Nancy Hôpital central
Nancy, France
Satisfaction questionnaire
The listener has to rate his appreciation of the Neuro 1 and Neuro 2 device and the sound quality on 7 points scale. The outcome measure corresponds to averaged score. The questionnaire has been developed by Oticon Medical to investigate satisfaction patient for their current and future speech processors. This questionnaire is not validated in the literature.
Time frame: 15 minutes
Speech comprehension in quiet
The speech material corresponds to the lists of Lafon designed for cochlear implant patients. Each lists contains 17 monosyllabic words. Speech comprehension is evaluated by the percentage of words and the percentage of phonemes correctly repeated. This double scoring is used to avoid floor or ceiling effects that could happen with only one of these measures.
Time frame: 15 mins
Speech comprehension in Noise
The speech material corresponds to the lists of Lafon designed for cochlear implant patients. Each lists contains 17 monosyllabic words. Speech comprehension is evaluated by the percentage of words and the percentage of phonemes correctly repeated. This double scoring is used to avoid floor or ceiling effects that could happen with only one of these measures.
Time frame: 15 mins
French Quick Sin
This test measures the SNR (Signal to Noise Ratio) at which the listeners understand half of the words of a sentence correctly. The outcome measure is in dB.
Time frame: 5 mins
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Speech intelligibility in noise measured with the test VRB (Vocale Rapide dans le Bruit, French version of the Quick Sin). This test measures the SNR (Signal to Noise Ratio) at which the listeners understand half of the words of a sentence correctly. The outcome measure is in dB. 3 lists are presented. Speech signals are presented at 60 dB SPL.