The aim of this descriptive study is to collect and describe the characteristics of objective measures to explore the specificity of algorithms.
Study Type
OBSERVATIONAL
Enrollment
153
Objective measures to be conducted with Custom SoundTM EP 5 software and the CP900 series sound processor in CI532 and CI512 Cochlear Implant recipients
HEARing CRC
Carlton, Australia
Hals-Nasen-Ohren-Klinik Universitätsklinikum Erlangen
Erlangen, Germany
Klinik für Hals-, Nasen-, Ohrenheilkunde Universitätsklinikum Frankfurt Goethe-Universität
Frankfurt am Main, Germany
Universitätsklinikum Schleswig- Holstein - Campus Kiel Klinik für Hals-, Nasen-, Ohrenheilkunde, Kopf- und Halschirurgie
Kiel, Germany
Specificity (%) of Transimpedance Measurement to Identify Electrode Fold-over
Measured with Custom SoundTM EP 5. It took one minute to measure all electrodes in surgery. Intraoperative transimpedance measures from 148 ears without fold-overs (as evidenced by radiographic imaging) were analysed to assess the specificity of the algorithm.
Time frame: at surgery, within one minute