The purpose of the DIAMOND-AF study is to establish the safety and effectiveness of the DiamondTemp System for the treatment of drug refractory, recurrent, symptomatic paroxysmal atrial fibrillation in patients.
The DIAMOND-AF study is a prospective, single blind, 1:1 randomized controlled study being performed at multiple centers in the United States, Canada and Europe. The study will evaluate the safety and effectiveness of the DiamondTemp System used for ablation in patients with paroxysmal atrial fibrillation (AF). Subjects will be randomized for treatment with either the DiamondTemp Ablation Catheter or the TactiCath™ Quartz Contact Force Ablation Catheter manufactured by Abbott. Patients will be followed for 12 months.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
SINGLE
Enrollment
482
a pulmonary vein isolation procedure will be performed using radiofrequency ablation catheter
a pulmonary vein isolation procedure will be performed using radiofrequency ablation catheter
Grandview Medical Center
Birmingham, Alabama, United States
Safety: Freedom From a Composite of Pre-specified Serious Adverse Events (SAEs)
The primary safety endpoint is defined as freedom from a composite of serious adverse events (SAE) occurring within 30-days and clinically symptomatic pulmonary vein stenosis through 6-months post-index ablation procedure, as adjudicated by an independent Clinical Events Committee (CEC) for relatedness to the procedure or device. The primary safety device- or procedure-related SAE composite will be the combined rate of the following events: * Atrioesophageal fistula * Bleeding complication * Cardiac tamponade / perforation * Death * Extended hospitalization * Myocardial infarction * Pericarditis * Phrenic nerve paralysis * Pulmonary edema * Pulmonary vein stenosis * Stroke post-ablation * Thromboembolism * Transient ischemic attack (TIA) post-ablation * Vagal nerve injury * Vascular access complications
Time frame: Within 30-days or 6-months after index ablation procedure
Effectiveness: Freedom From Documented Atrial Fibrillation(AF), Atrial Flutter(AFL) and Atrial Tachycardia(AT) Episodes Following the Blanking Period (3M Post-ablation) Through the End of the Effectiveness Evaluation Period (12M Post-ablation).
The primary effectiveness failure is defined by any of the following events: * Inability to electrically isolate all accessible targeted pulmonary veins during the ablation procedure * Documented episodes of AF, AFL or AT lasting ≥ 30 seconds in duration as evidenced by electrocardiographic data during the effectiveness evaluation period * DC cardioversion for AF, AFL or AT during the effectiveness evaluation period * A repeat ablation procedure to treat AF, AFL or AT during the effectiveness evaluation period * Use of a new or modification to existing Class I-IV anti-arrhythmic drug (AAD) regimen to treat AF, AFL or AT recurrence during the effectiveness evaluation period * Use of a non-study device for ablation of any AF targets during the index or repeat ablation procedure during the blanking period * More than one (1) repeat ablation procedure during the blanking period
Time frame: 3-12M (3-12 months) after index ablation procedure
Mean Duration of Individual Radiofrequency (RF) Ablations (Seconds)
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
University Of Alabama
Birmingham, Alabama, United States
Keck School Of Medicine
Los Angeles, California, United States
Sequoia Hospital
Redwood City, California, United States
Florida Hospital Orlando
Orlando, Florida, United States
Ochsner Medical Center
New Orleans, Louisiana, United States
Massachusetts General Hospital
Boston, Massachusetts, United States
Jackson Heart Clinic
Jackson, Mississippi, United States
Icahn School of Medicine at Mount Sinai
New York, New York, United States
Montefiore Medical Center
New York, New York, United States
...and 13 more locations
Mean duration of individual RF ablations (seconds) during the index ablation procedure
Time frame: Index ablation procedure
Mean Cumulative RF Time Per Procedure (Minutes)
Mean cumulative RF time per procedure (minutes) during the index ablation procedure
Time frame: Index ablation procedure
Freedom From a Composite of SAE Occurring Within 7-days
Freedom from a composite of SAE occurring within 7-days post-index ablation procedure as adjudicated by an independent CEC for relatedness to the procedure or device. The device- or procedure-related SAE composite will be the combined rate of the following events: * Atrioesophageal fistula * Bleeding complication * Cardiac tamponade / perforation * Death * Extended hospitalization * Myocardial infarction * Pericarditis * Phrenic nerve paralysis * Pulmonary edema * Pulmonary vein stenosis * Stroke post-ablation * Thromboembolism * Transient ischemic attack (TIA) post-ablation * Vagal nerve injury * Vascular access complications
Time frame: Within 7-days after the index ablation procedure
Freedom From Documented AF, AT and AFL Episodes in the Absence of Class I and III Anti-arrhythmic Drugs (AADs).
Freedom from documented AF, AT and AFL episodes following the blanking period through 12-month follow-up post-ablation procedure in the absence of class I and III anti-arrhythmic drug therapy.
Time frame: 3-12 months after index ablation procedure
Rate of Acute Procedural Success
Rate of acute procedural success is defined as confirmation of electrical isolation of PVs via assessment of entrance block at least 20 minutes following the last ablation around the respective PV.
Time frame: Index ablation procedure
Rate of Single Procedure Success With Freedom From Documented AF, AT and AFL at 12 Months.
Rate of single procedure success is defined as the rate of subjects treated with one single ablation procedure during study participation and with freedom from documented AF, AT and AFL at 12 months.
Time frame: Index ablation procedure through 12-months after index ablation procedure
Rate of Single Procedure Success With Freedom From ALL Primary Effectiveness Endpoint Failure Criteria.
Rate of single procedure success is defined as the rate of subjects treated with one single ablation procedure during study participation and with freedom from ALL primary effectiveness endpoint failure criteria.
Time frame: Index ablation procedure through 12-months after index ablation procedure
Rate of Occurrence of Electrically Reconnected Pulmonary Veins (PVs)
Rate of occurrence of electrically reconnected PVs following a 20-minute waiting period assessed by entrance block at index procedure.
Time frame: Index ablation procedure
Accumulated Changes in Quality of Life (QOL) Using the Atrial Fibrillation Effect on QualiTy-of-Life (AFEQT) Questionnaire
Accumulated changes in QOL using the AF QOL Survey (AFEQT Questionnaire) from baseline through 6 and 12 months following ablation procedure. The Atrial Fibrillation Effect on QualiTy-of-life (AFEQT) Questionnaire is an atrial fibrillation-specific health-related quality of life Questionnaire. The overall AFEQT score can range from 0 to 100, with 0 corresponding to complete disability and 100 corresponding to no disability. So a higher AFEQT score means a better outcome.
Time frame: Baseline, 6-months after index ablation and 12-months after index ablation
Neurological Changes Measured Using the National Institutes of Health Stroke Scale (NIHSS)
Neurological changes measured using the NIH stroke scale between baseline and post-ablation (pre-discharge visit) and at 12 months post-ablation procedure. The National Institutes of Health Stroke Scale (NIHSS) is a systematic assessment tool that provides a quantitative measure of stroke-related neurologic deficit. The total NIHSS score can range from 0 to 42, with 0 indicating no stroke symptoms and 42 indicating extremely severe stroke symptoms. So a higher NIHSS score means a worse outcome.
Time frame: Baseline, pre-discharge after index ablation and 12-months after index ablation procedure
Total Procedure Time (Minutes)
Total procedure time (minutes) at index procedure is defined as time of first assigned ablation catheter insertion into the vasculature to time of last procedural ablation catheter removed.
Time frame: Index ablation procedure
Time to Achieve Initial Pulmonary Vein Isolation (PVI) (Minutes)
Time to achieve initial PVI (minutes) at index procedure is defined as time of delivery of first RF ablation with the assigned ablation catheter until confirmation of PVI.
Time frame: Index ablation procedure
Total Treatment Device Time (Minutes)
Total treatment device time (minutes) at index procedure is defined as time of delivery of first RF ablation with the assigned ablation treatment catheter to removal of the treatment catheter.
Time frame: Index ablation procedure
Total Number of RF Ablations Per Procedure
Total number of RF ablations per procedure at index procedure
Time frame: Index ablation procedure
Total Fluid Infused Through the Ablation Catheter (mL)
Total fluid infused through the assigned ablation catheter (mL) at index procedure
Time frame: Index ablation procedure
Total Fluoroscopy Time (Minutes)
Total fluoroscopy time (minutes) at index procedure
Time frame: Index ablation procedure
Number of Re-hospitalizations Due to Atrial Fibrillation Recurrence After Blanking Period
Number of re-hospitalizations due to atrial fibrillation recurrence after blanking period
Time frame: 3-12 months after index ablation procedure