The study procedure of simultaneous VFSS and DDS measurement will be completed in one day and the subject will be followed within 2 business days after the study procedure to monitor for adverse events.
DDS signals and VFSS will be recorded simultaneously (for the same bolus) using barium contrast agent stimuli prepared in three consistencies: thin, mildly-thick and moderately-thick. Subjects will undergo VFSS with simultaneous DDS using up to 5 boluses of thin barium stimulus ("THIN-Ba"), and up to 4 boluses of barium thickened to mildly ("MILD-Ba") thick and up to 4 boluses of moderately ("MODERATE-Ba") thick barium consistencies using Resource Thicken Up Clear Nestlé Health Science (TUC). 4, 3 and 3 boluses for THIN-Ba, MILD-Ba and MOD-Ba will be analyzed using the classifier algorithms for sensitivity/specificity results. According to the exploratory trial, VFSS data for safety or efficiency can be missing for up to 14% boluses due to quality of VFSS recording. To compensate for potential losses of boluses due to missing gold standard (VFSS) data, 5, 4 and 4 boluses will be collected for the three consistencies respectively . The DDS signals will be sent to a dedicated application software installed at the CRO, which interprets the acceleration data and displays the examination result. The VFSS recording will be sent to CRO and provided for blinded assessment by the independent central VFSS laboratory.The study procedure of simultaneous VFSS and DDS measurement will be completed in one day and the subject will be followed within 2 business days after the study procedure to monitor for adverse events.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
SCREENING
Masking
NONE
Enrollment
452
The Nestle Health Sciences (NHSc) Dysphagia Detection System (DDS) is a portable, non-invasive device designed for use at the bedside. The investigational DDS has 3 basic components: Sensor Unit (suspended on a necklace), Sensor Fixation and a personal computer (PC) for collecting data. The Sensor Unit consists of a dual-axis accelerometer in a plastic housing that is attached to the front of a patient's neck just below the thyroid cartilage by the single-use, disposable fixation unit. The Sensor Unit is connected via a cable to an A/D converter which then connects via cable to the PC. The PC collects the examination data, which is then sent to dedicated application software (installed at the CRO), which interprets the acceleration data and displays the examination result.
Rancho Research Institute, Rancho Los Amigos National Rehabilitation Center
Downey, California, United States
University of Colorado Denver
Aurora, Colorado, United States
Medstar Rehabilitation Hospital
Area Under the Receiver Operating Characteristic (ROC) Curve to Compare the DDS Predicted Swallow Safety Outcome (Binary) vs. Clinical Reference Standard VFS for THIN-Ba
The primary efficacy of the DDS was measured as the sensitivity \& specificity obtained from comparing the DDS predicted swallow safety outcome (binary) with the clinical reference standard VFSS swallow safety outcome (binary) for thin barium (THIN-Ba) stimuli, using 5 boluses per subject protocol. ROC (AUC) was used for comparing the 2 algorithms. The ROC curve incorporates both sensitivity (true positive rate) and specificity (true negative rate) providing a single assessment incorporating both measures. The higher the total area under the curve, the greater the predictive power of the outcome. Primary analysis for futility was based on 242 subjects (Interim Analysis). The study was terminated based on futility as the AUC for primary endpoint (0.64) was less than the guiding futility bound of 0.75.
Time frame: The study procedure of simultaneous VFSS and DDS measurement using thin barium (THIN-Ba) was completed in one day for each subject.
AUC Under ROC Curve to Compare the DDS Predicted Swallow Safety Outcome (Binary) vs. Clinical Reference Standard VFS for MILD-Ba
The sensitivity \& specificity obtained from comparing the DDS predicted swallow safety outcome (binary) with the clinical reference standard VFSS swallow safety outcome (binary) for MILD barium (MILD-Ba) stimuli, using 4 boluses per subject protocol. ROC (AUC) was used for comparing the 2 algorithms. The higher the total AUC, the greater the predictive power of the outcome.
Time frame: The study procedure of simultaneous VFSS and DDS measurement using mild barium (MILD-Ba) was completed in one day for each subject.
AUC Under ROC Curve to Compare the DDS Predicted Swallow Safety Outcome (Binary) vs. Clinical Reference Standard VFS for MOD-Ba
The sensitivity \& specificity obtained from comparing the DDS predicted swallow safety outcome (binary) with the clinical reference standard VFSS swallow safety outcome (binary) for MODERATE barium (MOD-Ba) stimuli, using 3 boluses per subject protocol. ROC (AUC) was used for comparing the 2 algorithms. The higher the total AUC, the greater the predictive power of the outcome.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Washington D.C., District of Columbia, United States
Shirley Ryan AbilityLab
Chicago, Illinois, United States
Marionjoy Rehabilitation Hospital
Wheaton, Illinois, United States
Kentucky Clinic
Lexington, Kentucky, United States
Boston Medical Center
Boston, Massachusetts, United States
Henry Ford Health System
Detroit, Michigan, United States
New York Presbyterian/Weill Cornell Medical Center
New York, New York, United States
New York Presbyterian Hospital/Columbia University Medical Center
New York, New York, United States
...and 3 more locations
Time frame: The study procedure of simultaneous VFSS and DDS measurement using MOD-Ba was completed in one day for each subject.
The Sensitivity & Specificity for Swallow Efficiency Using THIN-Ba
The sensitivity \& specificity obtained from comparing the DDS predicted swallow efficiency outcome (binary) with the clinical reference standard VFSS swallow efficiency outcome (binary) for Thin barium (THIN-Ba) stimuli.
Time frame: The study procedure of simultaneous VFSS and DDS measurement using THIN-Ba was completed in one day for each subject.
The Sensitivity & Specificity for Swallow Efficiency Using MILD-Ba
The sensitivity \& specificity obtained from comparing the DDS predicted swallow efficiency outcome (binary) with the clinical reference standard VFSS swallow efficiency outcome (binary) for MILD-Ba.
Time frame: The study procedure of simultaneous VFSS and DDS measurement using MILD-Ba was completed in one day for each subject.
The Sensitivity & Specificity for Swallow Efficiency Using MOD-Ba
The sensitivity \& specificity obtained from comparing the DDS predicted swallow efficiency outcome (binary) with the clinical reference standard VFSS swallow efficiency outcome (binary) for MOD-Ba stimuli.
Time frame: The study procedure of simultaneous VFSS and DDS measurement using MOD-Ba was completed in one day for each subject.