The novel Coaxial Optical Coherence Tomography (OCT) System is designed to visualize both structures of the retina and choroid. The purpose of this research is the safety and evaluation of a combined coaxial optical coherence tomography (OCT) system, to image diseases of the posterior and anterior segment of the eye.
Background and Rationale: Many posterior segment ocular diseases involve the retinal and choroidal vasculature. A new technology to noninvasively and simultaneously visualize vascular pathology is of utmost importance to reduce patient exposure to different diagnosis methods like e.g. fluorescence angiography. The novel Coaxial Optical Coherence Tomography (OCT) System is designed to visualize both structures of the retina and choroid. Furthermore, it may be used to image other structures of the eye. Objective(s): The purpose of this research is the safety and evaluation of a combined coaxial optical coherence tomography (OCT) system, to image diseases of the posterior and anterior segment of the eye. This device may be useful for the early diagnosis and monitoring of a variety of diseases involving the eye, such as Glaucoma, Diabetic Retinopathy or Age-Related Macular Degeneration. Measurements are compared to normal controls. Measurements and procedures: After screening, all participants will undergo a scanning procedure on the same day. Only one eye of each participant will be scanned with the investigational device. For patients, the eye showing the disease will be chosen, for healthy volunteers, the left or right eye will be scanned according to a randomisation list. The image quality is rated by the investigator and an independent assessor on a VAS (visual analogue scale). A subgroup of each stratum will have a comparative measurement of the choroid thickness against a SS-OCT (swept-source-OCT) device (Topcon) and a retina thickness against the Spectralis data. The subgroup will consist of half planned sample size. The measurement is taken from one eye, ie. every second participant per stratum will be attributed to the thickness measurement subgroup. The endpoint is a true/false criterion on whether the measurement of the retina and choroid thickness was successful. The thickness is measured in μm. The measurement is considered a "success" if the value is no more than +/- 10% off the reference measurement.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
30
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
Visual acuity (VA) after investigation
Safety of the imaging investigation expressed as visual acuity before (baseline) and after imaging, and tolerability assessment questionnaire.
Time frame: Within 1 hour after investigation
Scan quality
Comparing scan quality of investigational device to scan quality of a reference device, expressed as thickness of the retina and the thickness of the choroid
Time frame: After completion of all measurement, an average of 2 years.
Procedural success
Procedural success that is defined as: Image Quality of anterior and posterior eye structures
Time frame: After completion of all measurement, an average of 2 years.
Procedural safety and comfort assessed by questionnaire
Procedural comfort in the opinion of the patients assessed with a questionnaire
Time frame: Within 1 hour after investigation