The study will assess the CT-800 and TRK-2P for conformity to ANSI Z80.10-2014 Ophthalmic Instruments - Tonometers, to the FDA Guidance for Industry and FDA Staff, Tonometer-Premarket Notification \[510(k)\] Submissions, and to the applicable Supplemental Information Sheet. This clinical device study will also assess the substantial equivalence of TRK-2P to a predicate device in regard to pachymetry.
Study Type
OBSERVATIONAL
Enrollment
345
tonometer
tonometer, pachymeter
Andover Eye Associates
Andover, Massachusetts, United States
Agreement of IOP
Agreement of the measured IOP between the test devices and the predicate device for the tonometer function of CT-800 and TRK-2P. Presentation of results and analyses will include descriptive statistics (mean, standard deviation, limits of agreement), Bland-Altman plot and linear regression.
Time frame: 1 day
Agreement of Corneal Thickness
Agreement of measured corneal thickness between the test device and the predicate device for the pachymeter function of the TRK-2P. Presentation of results and analyses will include descriptive statistics (mean, standard deviation, limits of agreement), Bland-Altman plot and Deming regression.
Time frame: 1 day