The purpose of this study is the verification of Cloud-DX Vitaliti™ Continuous Automated Non-Invasive Blood Pressure Monitor for Clinical Use and Self Measurement according to standards laid out by IEEE Std 1708 and AAMI-ISO 81060-2.
Several automated vital sign measurement systems have been developed recently; however, non-invasive continuous blood pressure monitoring continues to be a major limitation. To circumvent the limitations of current systems of interval hemodynamic measurement and increasing demands on health care providers, a non-invasive automated continuous vital sign monitor that can integrate into hospital early warning systems and warn health care providers of deteriorating vital sign parameters would be of significant value and a major advance for current hospital monitoring systems. Cloud-DX Vitaliti™ is a new low profile automated vital sign monitor, which includes a continuous non-invasive blood pressure measurement system. Additional parameters are also measured. Paired with companion software from cloud based computing, Vitaliti™ features a user friendly interface, onboard radios for communication with smartphones/tablets/personal computers, and several sensors. Cloud-DX Vitaliti™ has undergone preliminary testing and is now ready for verification to achieve compliance with AAMI/ISO and IEEE standards of medical grade device. At the Hamilton General Hospital (HGH), Cardiac Surgical Intensive Care Unit (CICU) and Vascular Surgical Step Down Units, arterial catheters are standard of care and performed on most patients for variable periods of post-operative recovery. Within this setting, the IEEE Std 1708 and AAMI-ISO 81060-2 would be fully met. Furthermore, the cardiovascular surgical caseloads at HGH would provide sufficient test subjects to validate VitalitiTM in a timely fashion.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
78
Measurements recorded by the Vitaliti device will be compared to the recordings from the arterial catheter.
Hamilton General Hospital
Hamilton, Ontario, Canada
Blood pressure
Determining the accuracy of non-invasive blood pressure measurement compared to invasive blood pressure. A detailed analysis will be conducted on the invasive blood pressure to verify accuracy as defined in the ISO 81060-2 Standard.
Time frame: 30 minutes