This study evaluates the performance of eight (8) cleared capnography carbon dioxide (CO2) sampling filterlines during patient simulated scripted activities when connected to a Medtronic (MDT) dual parameter monitor.
The eight (8) Capnography CO2 Sampling Filterlines (CCSF) have different designs for obtaining a gas sample for CO2 measurement and also for delivering supplemental oxygen (O2). In this study awake, healthy volunteers will evaluate the 8 CCSF (Capnography CO2 Sampling Filterline) to determine their performance variability in providing a representative exhaled CO2 sample for measurement on the Capnostream 35 dual parameter monitor during scripted activities of closed mouth/open mouth breathing during both high and low respiration rate with supplemental oxygen delivery at 5lpm.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
30
Use of 8 post-market nasal CO2 cannula sampling filterlines; 4 nasal only sampling ports manufactured by Medtronic, Salter, Hudson/Teleflex and FlexiCare) and 4 post-market oral/nasal sampling ports manufactured by Medtronic, Salter, Dispo-Med and Westmed during scripted activity with supplemental O2.
Clinimark LLC
Louisville, Colorado, United States
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
Comparison of measured end-tidal CO2 (EtCO2), Rebreathing as measured by fractional inspiratory CO2 (FiCO2) as a function the activity for 8 different CCSF designs tested on 23 subjects with supplemental O2 at 5lmp for four activities. Device data collected a 20Hz (20 data points/second) for the 3 hours of enrollment. Participants lateral lying with head on a pillow with supplemental O2 during each nasal only or open mouth breathing at respiration rate of 6 and 24.
Time frame: Three hours
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
Gas sample provided by CCSF to the Capnostream 35 for measurement of the partial pressure of CO2 as a function of time. The Capnostream 35 Respiration rate algorithm counts the breaths per minute based upon the frequency of a CO2 validated breath waveform.
Time frame: Three hours
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
Gas sample provided by CCSF to the Capnostream 35 for measurement of the partial pressure of CO2 as a function of time. The Capnostream 35 Respiration rate algorithm counts the breaths per minute based upon the frequency of a CO2 validated breath waveform.
Time frame: Three hours