The investigator's objective is to validate the novel automated blood pressure device against the requirements set forth in the AAMI\_ISO 81060-2:2010 standard in voluntarily consented study participants.
A new prototype blood pressure cuff has been designed for the higi SH llc health stations. This new cuff accurately captures blood pressure (BP) by emulating a fitted cuff which is used in conjunction with a mercury sphygmomanometer, the accepted gold standard for measuring BP values in most clinical facilities. The cuff is driven by a motor and gearhead to automatically "wrap" the cuff around a user's arm when they initiate a BP test. A BP measurement is then taken by an Original Equipment Manufacturer (OEM) BP module using an algorithm developed for use with a fitted cuff; the same module one would expect to find in a clinic or doctor's office. For the device to pass the AAMI\_ISO 81060-2:2010 standard statistical requirements and thus, be considered a valid device for measuring BP it will need to meet the standards criteria 1 and 2.
Study Type
OBSERVATIONAL
Enrollment
85
Validation of an automated blood pressure device
University of Southern California
Los Angeles, California, United States
Systolic and Diastolic blood pressure (criteria 1 and 2)
Mean value of the differences of the paired (i.e., device under test - reference method) determinations for all study participants shall be within or equal to ± 5 mmHg with a standard deviation no greater than 8 mmHg (criterion 1). Standard deviation of the averaged paired determinations per participant shall meet the criterion 2 listed in the standard.
Time frame: 60 minutes