This is a multi-center study with a minimum of three sites in the United States. The study will enroll approximately 1750 female subjects and will have a study duration of approximately 9 months after enrollment of the first subject. Female subjects seen at the participating sites for any reason will be evaluated for enrollment in this study. All subjects will be managed per standard of care as applicable. Subjects who are enrolled in the study will perform self-collection of a vaginal swab to be tested by Click device, and allow the health care provider (HCP) to collect three additional vaginal swabs to be tested by recognized FDA-cleared comparator methods. Subjects will complete the study in a single visit. The primary objective is to assess the performance of the Click device for detection of Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Trichomonas vaginalis (TV) in self-collected vaginal specimens as compared to Patient Infected Status (PIS) determined by three approved comparator assays using vaginal specimens collected by a qualified HCP in support of obtaining FDA clearance.
This study is a multi-center study with a minimum of three Clinical Laboratory Improvement Amendments-waived (CLIA-waived) intended operator sites in the United States; approximately one-third of the total number of sites will have a patient population of low Neisseria gonorrhoeae (NG), Chlamydia trachomatis (CT), or Trichomonas vaginalis (TV) prevalence. Low prevalence sites are defined as sites with a prevalence \< / = 2% for any of the three targets. Sites with a prevalence higher than 2% for any of the three targets will be defined as high prevalence sites. The study will enroll approximately 1750 female subjects and will have a study duration of approximately 9 months after enrollment of the first subject. Approximately one-third of the total subjects shall be recruited from low prevalence sites for the infections of interest. Female subjects seen at the participating sites for any reason will be evaluated for enrollment in this study. All subjects will be managed per standard of care as applicable. Subjects who are enrolled in the study will perform self-collection of a vaginal swab to be tested by Click device, and allow the health care provider (HCP) to collect three additional vaginal swabs to be tested by recognized FDA-cleared comparator methods. Subjects will complete the study in a single visit. The hypothesis of this study is that the identification of each organism (CT, NG, and TV) in self-collected vaginal swabs by women using the Click device will agree with the Patient Infected Status (PIS) with a high sensitivity and specificity. The primary objective is to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP in support of obtaining FDA clearance. The secondary objectives are: 1) to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens among symptomatic subjects as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP; 2) to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens among asymptomatic subjects as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
295
A single-use device containing a polymerase chain reaction (PCR)-based nucleic acid amplification test (NAAT) for detection of deoxyribonucleic acid (DNA) from Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Trichomonas vaginalis (TV) in self-collected vaginal specimens.
Axis Clinical Trials
Los Angeles, California, United States
University of California, San Diego - Antiviral Research Center
San Diego, California, United States
San Francisco Department of Public Health - San Francisco City Clinic
San Francisco, California, United States
Florida International University - Student Health Center
Miami, Florida, United States
Cook County Health and Hospitals System - Ruth M Rothstein CORE Center
Chicago, Illinois, United States
Johns Hopkins Hospital - Medicine - Infectious Diseases
Baltimore, Maryland, United States
University of Mississippi - Infectious Diseases
Jackson, Mississippi, United States
Magee Women's Hospital of UPMC - Reproductive Infectious Disease Research
Pittsburgh, Pennsylvania, United States
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCP
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1