The aim of this study is to collect capnometric (tidal breathing CO2) data using TidalSense's N-Tidal C Data Collector Device. The data collected will be examined for correlation with currently accepted assessment techniques for COPD (spirometry, vital signs and pulse oximetry) and the patient's clinical condition (stable, worsening, improving). The patient numbers and study duration have been determined by power calculations to provide a statistically significant sample size, based on expected number of exacerbations (mild, moderate and severe) from the patient cohort. The data will also be examined to assess the predictive possibility of a non-invasive self-care personal CO2 monitoring device to identify COPD exacerbations.
Study Design: This is an uncontrolled, non-randomised, un-blinded longitudinal observational data collection study with the N-Tidal C Data Collector Device in patients with COPD. Patients will be identified from the Cambridge COPD Centre. This unit sees patients following emergency admissions, GP referrals and has a regional referral base for complex COPD. Patients' medical records will be reviewed and classified according to GOLD criteria. Those patients who are classified as COPD with appropriate inclusion criteria (below) will be invited to participate. Patients who give informed consent will be screened for the study. The N-Tidal C Data Collector Device will be explained to the patients and they will be trained how to use it. All patients will complete a short daily diary record. Patients will undergo a clinical assessment of COPD during screening at Addenbrooke's Hospital. All patients will be assessed five times: at the start, 2 weeks, 10 weeks, 18 weeks and at the end of the 26 week study period. A brief follow-up telephone review will be conducted approximately 2 weeks after the end of the monitoring period. At each assessment, capnometry measurements will be taken in addition to vital signs and pulse oximetry. The study will not interfere with routine management of the patient's condition.
Study Type
OBSERVATIONAL
Enrollment
50
Addenbrooke's Hospital
Cambridge, Cambridgeshire, United Kingdom
To measure the changes in the α angle of the tidal breathing CO2 (TBCO2) waveform data from stable COPD through an exacerbation and back to stable.
The α angle is one of the key parameters extracted from the TBCO2 waveform data captured by the N-Tidal C device. The α angle is given in degrees and significant changes in the α angle have previously been observed between stable COPD and acute exacerbations of COPD.
Time frame: Throughout the 26 week use of the N-Tidal C device for each patient.
To measure the minimum and maximum CO2 levels of the TBCO2 waveform data from stable COPD through an exacerbation and back to stable.
The minimum and maximum CO2 levels are key parameters extracted from the TBCO2 waveform data captured by the N-Tidal C device. CO2 levels are given in kPa and significant differences have previously been observed between stable COPD and acute exacerbations of COPD.
Time frame: Throughout the 26 week use of the N-Tidal C device for each patient.
To measure the variability of breath length of the TBCO2 waveform data from stable COPD through an exacerbation and back to stable.
The variability of breath length is one of the key parameters extracted from the TBCO2 waveform data captured by the N-Tidal C device. Breath length is measured in seconds and significant differences have previously been observed between stable COPD and acute exacerbations of COPD.
Time frame: Throughout the 26 week use of the N-Tidal C device for each patient.
To measure the standard deviation of End Tidal CO2 levels of the TBCO2 waveform data from stable COPD through an exacerbation and back to stable.
The standard deviation of End Tidal CO2 levels is one of the key parameters extracted from the TBCO2 waveform data captured by the N-Tidal C device. End Tidal CO2 levels are given in kPa and significant differences have previously been observed between stable COPD and acute exacerbations of COPD.
Time frame: Throughout the 26 week use of the N-Tidal C device for each patient.
Longitudinal Database
To collect a longitudinal observational database of TBCO2 waveform records for up to 50 patients with COPD over a period of 26 weeks using the N-Tidal C device.
Time frame: Throughout the 26 week use of the N-Tidal C device for each patient.
COPD Exacerbations
To capture the carbon dioxide signature predictive of exacerbations in monitored patients who undergo mild, moderate and severe exacerbations during the study.
Time frame: Throughout the 26 week use of the N-Tidal C device for each patient.
COPD Variability
To identify the within-day and day-to-day variability of respired CO2 in patients with COPD
Time frame: Throughout the 26 week use of the N-Tidal C device for each patient.
Usability of the N-Tidal C device
Measured by the patients compliance and perseverance to the use the N-Tidal C device.
Time frame: Throughout the 26 week use of the N-Tidal C device for each patient.
Acceptability of the N-Tidal C device
The patients will be asked questions about the ease of use of the N-Tidal C and twice daily respiratory monitoring.
Time frame: At the end of the 26 week use of the N-Tidal C device for each patient.