Dual-energy X-ray Absorptiometry is frequently used to measure bone mineral density. A new medical device, Box 3D DXA, creates a 3D image using a statistics reconstruction model developed on the femur. This new imaging technique does not require additional irradiation and ought to improve bone measures as well as incorporating densitometric parameters into the diagnosis. This study will test the reconstruction of the 3D image from lombar spine measurements and compare accuracy against measurements made using CT scan.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
OTHER
Masking
NONE
Enrollment
120
Two dual-energy X-ray scans + CT scan
CHU Nimes
Nîmes, France
Concordance in lombar spine measurements between box 3D-DXA (index test) and CT scan (reference test)
Lin's concordance coefficient
Time frame: Day 0
Concordance in lombar spine measurements between box 3D-DXA (index test) and CT scan (reference test)
Bland Altman plot
Time frame: Day 0
Reproducibility of the 3D-DXA Spine algorithm for lombar spine measurements
2 acquisitions under the same conditions. Using root-mean-square coefficient of variation to calculate the Least Significant Change
Time frame: Day 0
Robustness of 3D-DXA Spine measurement for lombar spine measurements
2 acquisitions under the same conditions. Using interquartile range as a measure of dispersion of the variable
Time frame: Day 0
Concordance in femur measurements between box 3D-DXA (index test) and CT scan (reference test)
Lin's concordance coefficient
Time frame: Day 0
Concordance in femur measurements between box 3D-DXA (index test) and CT scan (reference test)
Bland Altman plot
Time frame: Day 0
Reproducibility of the 3D-DXA Femur algorithm for femur measurements
2 acquisitions under the same conditions. Using root-mean-square coefficient of variation to calculate the Least Significant Change
Time frame: Day 0
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Robustness of 3D-DXA Femur measurement for femur measurements
2 acquisitions under the same conditions. Using interquartile range as a measure of dispersion of the variable
Time frame: Day 0