The study will be an open label laboratory study with all subjects receiving HFCWO with The Monarch® System Objective: Assess device settings and to identify frequency/pressure (intensity) combinations that produce high airflow and oscillating volume Methodology: Subjects will receive HFCWO using The Monarch® System at multiple frequency / intensity combinations on a single visit day. Frequency / intensity combinations will be evaluated to determine which settings produce highest airflow and highest oscillating volume.
The study will be an open-label, laboratory study. All enrolled subjects will receive HFCWO therapy with the Monarch® system. The study will be conducted at one (1) site in the US. The study will evaluate oscillating airflow and oscillating volume induced during therapy with the Monarch® system. Subjects will also rate the comfort of the therapy. Patients with cystic fibrosis (CF) who meet all of the inclusion criteria and none of the exclusion criteria will be enrolled in the study. Therapy with the Monarch® system will be delivered using multiple frequency/intensity combination settings. The order of settings will be randomized for each study subject. During therapy oscillating flow and volume measurements will be obtained for each frequency/intensity combination. Following each frequency/intensity combination, subjects will respond to a subjective assessment of the comfort of the therapy by rating the comfort on a visual analogue scale (VAS).
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DEVICE_FEASIBILITY
Masking
NONE
Enrollment
10
HFCWO Airway Clearance Device
Peak oscillating expiratory airflow measurements
Subjects will be treated with the Monarch® system using multiple frequency/intensity combination settings. After adjusting the therapy to the desired setting, subjects will be instructed breath through the mouthpiece for the flow measurement system, to breathe normally and continue relaxed breathing. Oscillating flow and volume measurement will be captured by the system over a period of 30 to 60 seconds of relaxed breathing. Data from the measurement of oscillating flow for each frequency/intensity combination setting will be processed following the completion of the testing
Time frame: 4-6 hours
Oscillating volume
Oscillating volume measurement will be conducted concurrently with the measurement of flow. Subjects will be treated with the Monarch® system using multiple frequency/intensity combination settings as described above. Data from the measurement of oscillating volume for each frequency/intensity combination setting will be processed following the completion of the testing
Time frame: 4-6 hours
Patient comfort/discomfort evaluated by 100 mm visual analogue scale (VAS)
Following the completion of testing for each frequency/intensity combination setting, subjects will rate the level of comfort/discomfort of the setting on a VAS. A Visual Analogue Scale is a psychometric response scale which will be used as a measuring instrument for subjective comfort or discomfort. It is a measurement instrument that ranges across a continuum of values and used to measure the intensity or frequency of various symptoms. * The Visual Analogue Scale ranges from (0 to 100 mm). * Zero means very uncomfortable/intolerable and 100 means very comfortable. The subject will mark the line for each setting combination to indicate the level of comfort or discomfort. After the Study visit, the research staff will measure the mark on the line and document the value.
Time frame: 4-6 hours