In this prospective device comparison study with 58 healthy eyes, the Topcon CT-1® showed statistically significant elevated intraocular pressure measurements compared to the gold standard Goldmann application tonometry.
Introduction: Intraocular pressure (IOP) measurement is crucial in glaucoma diagnosis and treatment. We compared a non-contact air tonometer (Topcon CT-1®) with the gold standard Goldmann Applikation tonometer (GAT). Methods: In this prospective, mono centric device comparison study, only healthy eyes were enrolled. IOP was measured three times with the Topcon CT-1® followed by three times GAT. Mean IOP was calculated and compared with a paired t-test, a linear regression and a Bland-Altman analysis.
Study Type
OBSERVATIONAL
Enrollment
58
Pressure tip or air puff is applied on eye
Augenpraxis Centerpassage
Wettingen, Canton of Aargau, Switzerland
Comparison of GAT with air tonometry Topcon CT1
Difference in mmHg between these two devices
Time frame: 7 months