The purpose of this study is to demonstrate the performance and safety of the Guardian™ Sensor (3) over 170 hours (7 days) when inserted in the abdomen used in subjects age 14 - 75 years with the Guardian™ Connect System.
This study is a multi-center, prospective single-arm design without controls. Up to 72 subjects will be enrolled in order to have approximately 60 subjects complete the study. Three investigational centers in China will be used during this study. During the study, each subject will be randomly assigned to one day of the Yellow Springs Instrument (YSI™\*) frequent sample testing (FST) (Day 1, 3-5, or 7).Subjects will wear two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days (one which will be paired to the Guardian Connect app and the other will function as a glucose recorder.). The Guardian Sensor (3)s will be worn in the abdomen area and self-inserted by the subject on same side or opposite sides. On the evening prior to FST, subjects will be asked to fast for approximately 12 hours and adjust their insulin and medications according to routine care (for example as they would do for fasting lipid panel). Subjects may fast for shorter period of time based on investigator discretion. The subject should be in fasting status upon arrival hospital to start FST process. The feeding protocol may be modified based on investigator discretion. The duration of FST will be approximately 7 hours. During the study, subjects will continue with their current diabetes regimen independent of the study devices. Subjects will be instructed by the investigational center that they are not to use the study devices (except for the CONTOUR™\* study meter) for the management of their diabetes. The CONTOUR™\* study meter may be used for treatment decisions and calibration of Guardian™ Sensor (3).
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
72
Use two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days.
The Seventh Medical Center of PLA General Hospital
Beijing, China
Sir Run Run Shaw Hospital
Hangzhou, China
Shanghai Sixth People's Hospital
Shanghai, China
Percentage of Guardian™ Sensor (3) Values That Are Within 20% Agreement of YSI
Percentage of Guardian™ Sensor (3) Values that are within 20% agreement of Yellow Springs Instrument (YSI) plasma glucose values (±20 mg/dL (1.1 mmol/L) when reference blood glucose (YSI) less than or equal to (≤) 80 mg/dL (4.4 mmol/L) ) during YSI frequent sampling testing (FST) days.
Time frame: 168 Hours
Consensus Error Grid Analysis of Mean Rate of Paired Sensor Values and YSI Values in Zone A+B
A mean rate in Zone A+B of Consensus Error Grid between Guardian Sensor (3) values and YSI™\* plasma glucose values during YSI™\* FST days was evaluated. Consensus Error Grid (or Parkes error grid) compared the paired primary sensor and YSI reference glucose values. Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action", Zone B as "altered clinical action with little or no effect on clinical outcome". Ideal situation is 100% in Zone A + B.
Time frame: 168 Hours
Clarke Error Grid Analysis of Mean Rate of Paired Sensor Values and YSI Values in Zone A+B
A mean rate in Zone A+B of Clarke Error Grid between Guardian Sensor (3) values and YSI™\* plasma glucose values during YSI™\* FST days was evaluated. Clarke Error Grid compared the paired primary sensor and YSI reference glucose values. Zone A are "clinically accurate in that they would lead to clinically correct treatment decisions". Zone B "would lead to benign or no treatment". Ideal situation is 100% in Zone A + B.
Time frame: 168 Hours
Mean Absolute Relative Difference (MARD)
Mean absolute relative difference (MARD) between Guardian Sensor (3) values and YSI™\* plasma glucose values during YSI™\* FST days was evaluated. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple FST days were pooled together for reporting purpose.
Time frame: 168 Hours