The study is to evaluate the accuracy of a test device called "PAMG-1" to see if a pregnant women has ruptured membranes in comparison to standardly used testing methods.
The study is to evaluate the accuracy of the PAMG-1 immunoassay in comparison to standard diagnostic methods for premature preterm rupture of membranes including the sterile speculum exam and the amniodye test as the gold standard. The procedure includes the insertion of a swab into the subject's vaginal vault without a speculum for one minute. Once the swab is obtained, the treating physician will place the swab in a solvent vial for one minute. A developing strip is then placed in a vial and allowed to develop for five minutes. The results are read from the developing strip. Processing of this swab will be performed without further involvement of the study participant. The insertion of the swab is a one time procedure and is the only addition to any standard of care that would applied to the patient during their clinical evaluation at that time. Given the study group and both pregnant and non-pregnant controls, standard of care procedures will include a wide variety of obstetrical and/or gynecological (OB/Gyn) procedures including but not limited to a vaginal exams, sterile speculum exams (including pap smears), biopsies, and amniocentesis. The swab for the PAMG-1 immunoassay will be performed after other evaluations are performed. Medical records of the subject and (if applicable) her newborn infant will be reviewed for data pertaining to the study, including, but not limited to, maternal and fetal outcomes at the time of enrollment and after delivery and history of the present illness requiring an OB/Gyn evaluation. The study participants will not be asked to do anything in addition to what is expected for the swab and their clinical visit. The PAMG-1 immunoassay will be obtained during the subjects regularly scheduled clinic appointment or hospital visit. No follow appointments will be needed.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
162
The procedure includes the insertion of a PAMG-1 swab into the subject's vaginal vault without a speculum for one minute. Once the swab is obtained, the treating physician will place the swab in a solvent vial for one minute. A developing strip is then placed in a vial and allowed to develop for five minutes. The results are read from the developing strip.
University of Iowa
Iowa City, Iowa, United States
Number of Participants With PAMG-1 Test Results That Matched the Results of Gold Standard Testing
Number of participants with PAMG-1 test results that matched the results of gold standard testing
Time frame: The assessment of the accurance PAMG-1 is completed by the end of the first study visit; one day.