Multi-center, open label, prospective clinical study to establish the acute safety and performance of the Cryterion Cardiac Cryoablation System
This is a single-arm, open-label clinical study using a cryoablation balloon device to isolate pulmonary veins during an atrial fibrillation ablation procedure. Cryo is a scientifically proven energy source used to create a thermal injury during endocardial ablation. The investigational device uses current technology with minor design modifications and will be studied to validate similarities in the safety and performance parameters compared to literature.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
88
Cryoablation System that includes a Balloon Catheter, Circular Mapping Catheter, Steerable Sheath and Cryoablation Console. In combination, components of the System will be inserted into the left atrium and a thermal injury at the PV ostium will be created to electrically isolate the veins
Clinical Hospital Centre Split
Split, Croatia
Catharina Ziekenhuis
Eindhoven, Netherlands
Erasmus MC
Rotterdam, Netherlands
Auckland City Hospital
Auckland, New Zealand
Primary Safety Outcome: Percentage of Participants With Procedure Related Adverse Events OR Device Related Adverse Events OR Serious Adverse Device Effects
Percentage of participants with procedure related adverse events OR device related adverse events OR Serious Adverse device effects including: Death, Myocardial Infarction, Cardiac perforation/pericardial tamponade, Cerebral Infarct or systemic embolism, Major bleeding requiring transfusion of blood products, Mitral or tricuspid valvular damage, phrenic nerve damage causing persistent diaphragmatic paralysis, Symptomatic pulmonary vein stenosis, atrio-esophageal fistula, Air embolism leading to life-threatening event, other serious adverse device effects (SADEs)
Time frame: 12 months
Acute Procedural Success
Percentage of patients achieving pulmonary vein isolation defined as entrance and exit block for each of the targeted pulmonary veins following the last ablation application.
Time frame: Following the last cryo-application for each of the targeted veins, during the index procedure. Mean procedure time was 118 minutes.
12 Month Treatment Success - Percentage of Subjects Free From Arrhythmias at 12 Months
A treatment failure is defined as a subject being an acute procedural failure, having more than one repeat procedure during the 90-day blanking period or having a documented, symptomatic episode(s) of atrial fibrillation or atrial tachycardia, between 91 and 365 days post index procedure. Occurrence or recurrence of a right atrial arrhythmia is not considered a treatment failure post the 90-day blanking period
Time frame: 12 months
Secondary Safety Outcome: Number of Procedure-related or Device Related Adverse Events.
All procedure-related or device related adverse events with causal relationship
Time frame: 12 months
Percentage of Subjects With Isolation Immediately Post Ablation and With Reconnection During the Entrance/Exit Block Testing
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Percentage of subjects that present PV isolation immediately after ablation but show reconnection after 30 minutes during entrance/exit block testing. This outcome is only measured in subjects that were enrolled under the protocol that required a 30 minute waiting period between isolation and verification of isolation.