The aim of this study is to investigate the possibility of optimizing the performance of CRT-D in non-responding patients through utilization of cardiac strain speckle tracking
In approximately 30% of patients, cardiac resynchronization therapy (CRT) fails to lead to any improvement of the patients' status. In this study, an investigation of a possible method of optimization through speckle tracking of cardiac strains is attempted. Patients not responding to Cardiac Resynchronization Therapy-Defibrillators (CRT-D) after 3 months are randomly divided into control and intervention groups. Atrioventricular interval is adjusted so that E and A waves do not overlap and the interventricular interval subsequently optimized to yield maximum improvement of the sum of longitudinal+radial+circumferential strains. The left ventricular ejection fraction (LVEF) and NYHA (New York Heart Association Classification improvement 3 months after optimization are evaluated and use of other strain combinations assessed.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
50
The previously implanted, ineffective CRT-D is reprogrammed under supervision of Trans-Thoracic Echocardiography (TTE) to: 1. adjust the atrioventricular interval so that E and A waves do not overlap 2. the interventricular interval is subsequently optimized to yield maximum improvement of the sum of longitudinal+radial+circumferential strains.
This intervention relates to the determination of strains and atrioventricular interval and supplies data for reprogramming the ineffective CRT-D Transthoracic echocardiography performed to determine the left ventricular ejection fraction as an outcome measure is not considered intervention here (as it is also performed in the control group and that would cause an error cross-referencing.
University Hospital Ostrava
Ostrava, Czechia
Change in the left ventricular ejection fraction
Left ventricular ejection fraction measured by transthoracic echocardiography prior to optimizing and during a follow-up examination three month after optimizing
Time frame: 6 months from CRT-D implantation, 3 months from optimization
Change in NYHA Classification
NYHA (New York Heart Association Classification measured using the standard NYHA range I to IV - I being the best result, IV the worst) was evaluated using a standard NYHA questionnaire before optimization and 3 months after optimization during a follow-up examination
Time frame: 6 months from CRT-D implantation, 3 months from optimization