Prospective, multicenter, observational, comparative clinical trial on the equivalence of two different OrthoPilot® navigation system generations applied for computer-assisted total knee arthroplasty
The study is designed as a prospective, non-interventional, multicenter, comparative cohort study. The product under investigation will be used in routine clinical practice and according to the authorized Instructions for Use (IfU). Those data that are obtained in routine clinical use will be documented in the Case Report Form (CRF). The study design is based on a two-arms design. The patients for the both groups will be recruited consecutively, starting with group A. The study will prove the equivalence of the groups regarding the outcome of the postoperative lower limb alignment (mechanical leg axis) and any deviation \>3° from neutral alignment is defined as an outlier.
Study Type
OBSERVATIONAL
Enrollment
217
Computer Assisted Total Knee Replacement with the OrthoPilot FS101 and the software TKA 5.1
Computer Assisted Total Knee Replacement with the with the OrthoPilot Elite and the software TKA 6.0
Universitätsklinikum Carl Gustav Carus
Dresden, Germany
Uniklinikum Freiburg
Freiburg im Breisgau, Germany
Schönkliniken Klinikum Eilbek
Hamburg, Germany
Rate of Lower Limb Alignment inside the range of 0±3°
A postoperative lower limb alignment of the knee prosthesis in reference to the mechanical leg axis (3, 18) outside the range of ± 3° is defined as the outlier. The outlier rates will be used as the efficacy criterion for groups comparison
Time frame: three months postoperatively
Lower limb alignment of the knee prosthesis in reference to the mechanical leg axis
The outcome and function of a prosthesis is influenced by the postoperative alignment of the leg and the orientation in reference to the axis. Incorrect positioning can lead to an increased implant wear and functional limitations (23, 24). The aim of a postoperative alignment within 3° varus to 3° valgus will be assessed using long leg x-rays 3-4 months postoperatively.
Time frame: 3 months postoperatively
Improvement of Knee Society Score (KSS)
The KSS score is composed of two subgroups and has a maximum score of 200.
Time frame: preoperatively and three months postoperatively
Improvement of Oxford Knee Score (OKS)
A patient-reported outcome instrument which contains 12 questions on activities of daily living that assess function and residual pain in patients undergoing total knee replacement (TKR) arthroplasty. Oxford Knee Score has a maximum of 48 points indicating satisfactory knee function
Time frame: preoperatively and three months postoperatively
Improvement of Quality of Life (EQ-5D-5L)
The EQ-5D-5L essentially consists of 2 pages: the EQ-5D descriptive system and the EQ visual analogue scale (EQ VAS). The descriptive system comprises five dimensions: mobility, self-care, usual activities, pain/discomfort and anxiety/depression. Each dimension has 5 levels: no problems, slight problems, moderate problems, severe problems and extreme problems. This decision results in a 1-digit number that expresses the level selected for that dimension. The digits for the five dimensions can be combined into a 5-digit number that describes the patient's health state.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Time frame: preoperatively and three months postoperatively
Learning Curve
The first 10 cases of each surgeon shall be excluded from the main analysis. The results of the first 10 cases ("learning curve") shall be assessed separately and compared with the "10 plus" cases ("routine cases") as a secondary learning curve analysis. The potential learning curve will be assessed measuring the intraoperative surgery time in reference to the already performed procedures of each surgeon
Time frame: Intraoperatively
Intraoperative handling of soft- and hardware components
Aim of this study is also to document any upcoming problems or limitations the operating surgeon has with the new components of the OrthoPilot® Elite hardware system and the new software application TKA 6.0 for the implantation of primary total knee endoprosthesis. Intraoperatively documented results of the navigation system will be recorded and integrated into the final analysis. Potential problems or comments will be analysed descriptively.
Time frame: Intraoperatively
(Serious) Adverse Device Effects
As part of the follow-up examination, any upcoming (serious) adverse device events or effects related or not related to the product under investigation, will be documented in the dedicated Case Report Forms. Recorded complications will be categorized and analyzed in order to assess the safety of the OrthoPilot® navigation system
Time frame: up to 3 months postoperatively