The CNOGA Combo Glucometer (CoG) employs an invasive glucose meter using blood glucose teststrips and requiering a capillary blood sample obtained from a fingerstick and in addition, a non-invasive optical component for prediction of tissue glucose at the fingertip. In this study, participants will use both devices during meal tests and also at home during routine care. The results from the device will be compared to a standard reference method and will be used to determine the accuracy of the two device components.
The trial starts with a screening and training visit (V1). This is followed by an individual calibration period for the non-invasive device component by means of multiple parallel measurements with the invasive and the noninvasive device component. This will be done by the participants at home. Thereafter, the participants will come to the study site and will ingest a standardized meal (V2). Blood will be drawn for assessment of HbA1c and biochemical safety parameters. During the next 3 hours, 11 fingersticks will be made to obtain capillary blood samples for comparator readings with the two CoG components and with a laboratory reference method (YSI gucose analyzer). The patients will now continue to use the devices at home for three months, but are not allowed to use any CoG result for treatment decisions. In an interim visit after 6 weeks (V3), blood will be drawrn for HbA1c and biochemical safety parameters and three comparison measurements will be made with the CoG device components and the laboratory reference method. At the final visit after 12 weeks (V4), the procedures of the first meal test visit (V2) will be repeated, which conludes that study. In addition, the participants will be asked to complete a device satisfaction questionnaire at V2, V3, and V4.
Study Type
OBSERVATIONAL
Enrollment
40
Generation of non-invasive and invasive glucose readings for comparison with a standard laboratory reference glucose determination method (YSI glucose analyzer)
Mills-Penisnsula Medical Center
San Mateo, California, United States
CoG accuracy in measuring glucose levels (non-invasively) during a standardized meal experiment.
Comparison of non-invasive CoG device measurements (tissue glucose in mg/dL) with the results of a standard laboratory reference method (YSI glucose analyzer, blood glucose in mg/dL) by consensus-error grid analysis
Time frame: Experiment will be performed twice within 3 months
CoG accuracy in monitoring glucose levels (invasively) during a standardized meal experiment.
Comparison of CoG invasive device readings (blood glucose in mg/dL) with the results of a standard laboratory reference method (YSI glucose analyzer, blood glucose in mg/dL) by consensus-error grid analysis
Time frame: Experiment will be performed twice within 3 months
Impact of the CoG on long-term (three month) glycemic control (HbA1c using standard laboratory procedures) in patients with type 1 and 2 diabetes
Change in HbA1c from baseline after 6 weeks and 12 weeks
Time frame: 3 months
Treatment satisfaction questionnaire
Change in device satisfaction questionnaire score after 6 weeks and 12 weeks
Time frame: 3 months