Premature infants often receive respiratory support and supplemental oxygen for a prolonged period of time during their admission in the NICU. While maintaining the oxygen saturation within a narrow target range is important to prevent morbidity, manual oxygen titration can be very challenging. Automatic titration by a controller has been proven to be more effective. However, to date the performance of different controllers has not been compared. The proposed randomized crossover trial Comparing Oxygen Controllers in Preterm InfanTs (COCkPIT) is designed to compare the effect on time spent within target range. The results of this trial will help determining which algorithm is most successful in controlling oxygen, improve future developments in automated oxygen control and ultimately reduce the morbidity associated with hypoxemia and hyperoxemia.
Both hypoxemia as hyperoxemia can potentially be harmful to premature infants. Oxygen titration during respiratory support is vital to prevent these conditions but is very challenging. In the investigator's neonatal intensive care unit preterm infants routinely receive automatic oxygen titration performed by a controller. The currently used controllers are both proven to be more effective than manual titration, however which of the two controllers is most effective in keeping oxygen saturation within target range remains unclear. This randomized crossover trial tests tests both controllers within every study patient to determine which controller is most effective and thereby would hopefully reduce morbidity associated with hypoxemia and hyperoxemia the most. The primary outcome measure is the proportion of time spent within target range, each controller will be tested for 24 hours within the same study subject. This is excluding a 1-hour wash-out period after a change in ventilator. Eligible infants are randomized to start with either the Oxygenie algorithm or CLiO2 algorithm and will switch to the other study arm after 24 hours of measurement.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
SINGLE
Enrollment
15
Automated oxygen control for 24 hours by the CLiO2 algorithm, preceded by a 1 hour wash-out period
Automated oxygen control for 24 hours by the Oxygenie algorithm, in case of switch in ventilator preceded by a 1-hour wash-out period.
Leiden University Medical Center
Leiden, South Holland, Netherlands
Proportion of time with SpO2 spent within set target range
Total time SpO2 is within the set target range (91-95%) including time spent above target range when in room air (set FiO2 \< 0.22)
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Proportion of time with SpO2 spent above target range (SpO2 > 95%)
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Proportion of time with SpO2 spent below target range (SpO2 < 91%)
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Coefficient of SpO2 variation
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Time in hypoxemic SpO2 ranges
Time in SpO2 range 85-90%, 80-84% and \<80%
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Time in hyperoxemic SpO2 ranges
Time in SpO2 range 96-98% and \>98%
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Frequency of episodes of hypoxemia
Episodes in hypoxemia with SpO2 \< 85% for ≥ 30 seconds, for ≥ 60 seconds and hypoxemia \< 80% for ≥ 30 seconds and ≥ 60 seconds
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Frequency of episodes of hyperoxemia
Episodes in hyperoxemia with SpO2 ≥ 97% for ≥ 30 seconds, for ≥ 60 seconds and hyperoxemia ≥ 99% for ≥ 30 seconds and ≥ 60 seconds
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Frequency of episodes of bradycardia
Episodes where the heart rate \< 100 beats per minute for ≥ 10 seconds
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Frequency of FiO2 adjustments during automated control
FiO2 adjustments as made by the controller and by bedside staff as manual over-rides by the bedside staff as well
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Average oxygen exposure
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Coefficient of variation of FiO2
Time frame: Continuously from randomization until study completion. (maximum of 50 hours) Measurements during wash-out periods (up to 1 hour after ventilator switch) will be excluded.
Effects on nursing workload in relation to FiO2 adjustment
Number of manual FiO2 adjustments and as the effect on workload measured by a translated questionnaire adapted from the NASA TLX questionnaire, using a visual rating scale
Time frame: From randomization until study completion. (maximum of 50 hours) Manual adjustments during wash-out periods will be excluded.
Confidence of the bedside staff in the automated control system
Measured at the end of each nursing shift by a subset of questions out of the System Trust Scale (Jian et al. 2000) questionnaire. All questions are answered on a 1 (not at all) to 7 (extremely) scale.
Time frame: From randomization until study completion. (maximum of 50 hours)