The goal of this study is to take into account channel interaction in the cochlear implant (CI) fitting process. Designing a fast and accurate method of measurement would help to customize the fitting procedures to patients' specificities in order to maximize speech understanding in every day conditions especially in noise.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
BASIC_SCIENCE
Masking
NONE
Enrollment
28
Subjects will be asked to repeat lists of words played by speakers in a noisy environment generated by the same speakers. A percentage of correct recognition is calculated for each list.
Sounds of variable center frequency are used as maskers, and a tone with a fixed frequency at a fixed level is the target. What is tested here is the level of masker that just masks the target.
Channel interaction assessment using eABRs (auditory nerve responses). The method consists of two measurements: 1) eABR measurements with stimulation by a single electrode of the cochlear implant at 70% of the dynamic range (four different electrodes are tested); 2) Measurement of a single eABR with stimulation from all four electrodes at 70% of the dynamic range. A comparison of the eABRs obtained by these two measurements, indicates the electrical and neural interactions between the electrodes.
ENT and Cervico-Facial Surgery Department, Hôpital Edouard Herriot
Lyon, France
Percentage of syllable correctly recognized during speech audiometry test
Subjects will be asked to repeat lists of words played by speakers in a noisy environment generated by the same speakers. A percentage of correct recognition is calculated for each list.
Time frame: At inclusion
Sound masking threshold
Measurement of channel interaction by establishing psychophysical tuning curves in forward masking. Sounds of variable center frequency are used as a masker, and a tone with a fixed frequency at a fixed level is the target. What is tested here is the level of masker that just masks the target.
Time frame: At inclusion
Electrically-evoked Auditory Brainstem Response (eABR)
Channel interaction assessment using eABRs (auditory nerve responses). The method consists of two measurements: 1) eABR measurements with stimulation by a single electrode of the cochlear implant at 70% of the dynamic range (four different electrodes are tested); 2) Measurement of a single eABR with stimulation from all four electrodes at 70% of the dynamic range. A comparison of the eABRs obtained by these two measurements, indicates the electrical and neural interactions between the electrodes.
Time frame: Up to 12 months after inclusion