In nearly 25% of children under 3 months, the sweat test produces a quantity of sweat that does not meet international recommendations and is insufficient to allow reliable and reproducible biological analyzes in the sweat collected. In children between 3 and 12 months, this rate is about 10% when it should not exceed 5%. Insufficient amount of sweat prevents confirmation or reversal of the early diagnosis of cystic fibrosis and early treatment before irreversible complications of the disease. In this trial, a new support of sweat collection (Macroduct® Advanced Model 3710 Sweat Collection System, Wescor) will be tested with the goal to increase the amount of sweat collected during the sweat test, in comparison with the clinical routine method.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
57
for each subject, both methods will be tested in order to compare the efficacy of the new sweat collection method (Macroduct® Advanced Model 3710 SYS, Wescor) to the method currently used in clinical routine.
Cliniques Universitaires St Luc
Brussels, Belgium
chloride concentration
The volume and the chloride concentration will be analyzed for each subject.
Time frame: day 1
sweat
sweat collected with sweat collection method (Macroduct® Advanced) Model 3710 SYS, Wescor) compared with that currently used in clinical routine. The effectiveness criterion of the new Macroduct Advanced collector will be defined as follows: the new device tested will be considered effective if the insufficient amount of sweat levels obtained are smaller or equal to those recommended, i.e. if it is lower than 10% in children under 3 months and below 5% in children between 3 and 12 months. The insufficient amount of sweat rates will be compared between the two methods.
Time frame: day 1